PUTTING THE INSTRUMENT INTO OPERATION
Axio Vert.A1
Mounting Optional Components
Carl Zeiss
05/2012
431030-7044-001
61
3.3
Mounting Optional Components
Switch off the microscope and pull the power plug before mounting any optional components.
Be sure to put the relevant assemblies back into operational condition after mounting optional
components. (See Sections 3.1 to 1.1.)
3.3.1
Changing Diaphragms in the Condenser with Modulator Disk
The condenser does not have to be taken out of its mount when diaphragms are to be installed or
changed.
•
Loosen the quick-locking mechanism of the condenser and turn the condenser in such a way that the
mounting hole on the right-hand or left-hand side of the transmitted light illumination carrier
becomes accessible. Tighten the quick-locking mechanism.
•
Remove the sealing cap (Fig. 3-25/
5
) from the mounting hole.
Each of the condenser positions on the modulator disk is marked with a number (Fig. 3-25/
16
).
•
Screw the desired module (Ph, iHMC, PlasDIC or DIC, Fig. 3-25/
3
) onto the thread (Fig. 3-25/
2
) of the
mounting tool.
Fig. 3-25 Installing Diaphragms in the Modulator Disk of the Condenser