![Zeiss Axio Vert.A1 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/zeiss/axio-vert-a1/axio-vert-a1_operating-manual_927645023.webp)
INSTRUMENT DESCRIPTION
Axio Vert.A1
Technical Specifications
Carl Zeiss
05/2012
431030-7044-001
23
2.4
Technical Specifications
Dimensions
Dimensions (Width x Depth x Height)
Axio Vert.A1 ....................................................................................... 235 mm
*
x 560 mm x 560 mm
Axio Vert.A1 FL or A1 FL-LED ........................................................... 235 mm
*
x 560 mm
***
x 560 mm
Axio Vert.A1 MAT ......................................................................... 220 mm
**
x 560 mm
***
x 355 mm
*
Equipped with fixed stage without object guide
** Without mechanical stage; with 40 mm x 40 mm mechanical stage: Width: 295 mm
*** Depth: 560 mm, without external (reflected light) illuminators
The above measurements are maximum dimensions, which include space for cable and connector outlets. The
depth was measured at an interpupillary distance of 62 mm, with binocular parts lowered.
Transmitted light LED illumination results in the height of transmitted light stands being reduced by 29 mm.
Weight****
Axio Vert.A1 Stand ................................................................................................................ 10.5 kg
Axio Vert.A1 FL-LED Stand ...................................................................................................... 12.3 kg
Axio Vert.A1 FL Stand ............................................................................................................ 11.7 kg
Axio Vert.A1 MAT Stand ........................................................................................................ 10.3 kg
**** Actual weight of the stand without any attachments, such as stages, tubes, etc.
Ambient Conditions
Shipping (in Packaging):
Permissible ambient temperature ............................................................................... -40 °C to +70 °C
Storage:
Permissible ambient temperature ................................................................................ +5 °C to +40 °C
Maximum permissible humidity (without condensation) ................................................. 75 % at 35 °C
Operation:
Permissible ambient temperature ................................................................................ +5 °C to +35 °C
Maximum permissible humidity (without condensation) ............................................... 75 % at +35 °C
Maximum altitude for use ...................................................................................................... 2,000 m
Atmospheric pressure ......................................................................................... 800 hPa to 1,060 hPa
Pollution degree ............................................................................................................................... 2
Operating Specifications
Operating environment ................................................................................................... Enclosed spaces
Safety class .............................................................................................................................................I
Degree of protection ....................................................................................................................... IP 20
Electrical safety ..................................................................As specified in DIN EN 61010-1 (IEC 61010-1),
taking into account CSA and UL regulations
Overvoltage category ............................................................................................................................. II
Radio interference suppression .......................................................... As specified in EN 55011 for Class B
Noise immunity ....................................................................................... As specified in DIN EN 61326-1
AC line voltage range ................................................................................. 100 VAC to 240 VAC ±10 %
Line frequency .................................................................................................................. 50 Hz to 60 Hz
Maximum power consumption of internal power supply unit ......................................................... 80 VA