Lock the cams (Fig. 3-25/
1
) on the underside of the mounting tool into the recesses (Fig. 3-25/
4
) of
the contrast module.
•
Insert the contrast module into the desired modulator disk position, using the gripping component
(Fig. 3-25/
6
) of the mounting tool.
•
The two centering screws (Fig. 3-25/
11
) may first have to be turned back slightly, as appropriate.
These screws are accessible from the outside through holes (Fig. 3-25/
13
).
•
During insertion tilt the equipped mounting tool and push the chamfer (Fig. 3-25/
8
) of the contrast
module mount against the aluminium-colored pressure piece (Fig. 3-25/
14
). At the same time (in the
case of DIC or PlasDIC) align coarsely cam (Fig. 3-25/
10
) and recess (Fig. 3-25/
12
).
The slit diaphragm and the DIC prism are automatically centered via the hole (Fig. 3-25/
15
) and the
shoulder (Fig. 3-25/
9
).
•
For the DIC prism and the PlasDIC diaphragm, rotate the alignment unit (Fig. 3-25/
7
) of the mounting
tool to allow the cam (Fig. 3-25/
10
) to engage into the recess (Fig. 3-25/
12
).
•
Hold the alignment unit (Fig. 3-25/
7
) and turn the mounting tool (Fig. 3-25/
6
) out of the contrast
module.
•
Remove the mounting tool and reinsert the sealing cap into the mounting hole.
•
Turn the condenser into the desired working position.
When aligning Ph diaphragms and iHMC modules, you may first want to precenter the
diaphragm in relation to the hole diameter on the upper side, based on your own visual
judgment, before employing the auxiliary microscope to perform the exact centering when the
contrast module is in the light path.
After the diaphragm change, attach identification labels to the front of the condenser.
The alignment of the iHMC module is described in Section 4.11.6.
To remove contrast modules from the modulator disk, follow the same approach, but proceed
in reverse order.