PUTTING THE INSTRUMENT INTO OPERATION
Carl Zeiss
Mounting Optional Components
Axio Vert.A1
68
431030-7044-001
05/2012
3.3.6
Mounting the Aqua Stop
The aqua stop should be inserted to protect the
objectives and the nosepiece whenever you work
with liquid specimens.
•
Detach the microscope stage and screw out the
objectives.
•
Place the collection tray (Fig. 3-34/
7
) onto the
nosepiece carrier (Fig. 3-34/
9
) and fasten the
tray with two screws (Fig. 3-34/
1
).
•
Before doing so, remove the two cover screws
from the tapped hole.
•
Put the cover plate (Fig. 3-34/
8
) onto the
nosepiece.
•
Equip the nosepiece with the objectives to be
used.
•
Put a fitting bellow (Fig. 3-34/
2
) over each
objective.
Make sure that the bellow sits on the cover plate.
Bellows come in two sizes: Size 1 - small, Size 2 - large.
−
Use the
small-diameter bellow
(431716-0160-000) to protect objectives having a front diameter of
16 mm to 22.5 mm
.
−
Use the
large-diameter bellow
(431716-0170-000) to protect objectives having a front diameter of
27.5 mm to 34 mm
.
When attaching bellows, see to it that the upper edge does not form any collection tray.
Close unused objective holes with the protective caps supplied.
•
Plug the cut-to-size drain hose (Fig. 3-34/
3
) onto the drain outlet (Fig. 3-34/
6
). Push the other end of
the hose through the plug of the collection bottle (Fig. 3-34/
4
) in such a way that the hose projects
some 3 mm to 4 mm downward from the plug.
The length of the hose must be such that the drain channel of the collection tray does not
bend during the focusing procedure.
•
Put the plug firmly onto the collection bottle.
•
Glue to the stand the Velcro fastener (Fig. 3-34/
5
) supplied. Use the Velcro to fasten the collection
bottle onto the stand.
•
Reattach the microscope stage.
After an accident involving liquids, be sure to remove the microscope stage and use a lint-free cloth to
dab off any remaining drops of liquid. It is particularly important that the front of the objective be
cleaned in order to ensure continued full functionality of the objective.
For cleaning suggestions, please refer to the brochure entitled "The Clean Microscope".
Fig. 3-34
Mounting the Aqua Stop