zebris Medical GmbH
Canid Gait Technical Specifications and User Manual
Page 27/33
NOTE
Before connecting the measuring system to the power supply, com-
pare the nameplate specifications on the power supply unit and on
the treadmill regarding the mains voltage and mains frequency, with
the local characteristics. Connect only if they are compatible.
WARNING
Before connecting or using the measuring system, carry out a visual
check of the power supply unit, power supply cable and socket, as
well as the earthing contacts. Damaged power supply units, cables
or plug and socket devices are to be replaced immediately by an au-
thorized person.
6.2.2 Connection of the System
For connecting the treadmill to the power supply, please additionally observe the respective
instructions in the user manual provided by the treadmill manufacturer.
WARNING
The connection of the treadmill and the Canid Gait power supply unit
must be done at a separate wall socket. It is not permissible to use
extension cables and/or multiple sockets.
WARNING
Set all connection cables in a way that prevents patients, or persons
taking part in the measuring procedure from tripping over them or
damaging them mechanically. For this purpose, run the cables using
cable protections or if necessary fasten the cables with adhesive tape
to the floor.