6
Setup and Operation of the Canid Gait System
6.1
Positioning of the measuring system
NOTE
For setup, installation and safety related instructions of the treadmill
please refer to the user manual of the treadmill manufacturer.
WARNING
A safety zone of at least 2 m in length and 1 m in width must be kept
free directly behind the treadmill. No items may be left in this zone
(video camera, etc.).
NOTE
When using SYNCCam or SYNCLightCam, a distance of approxi-
mately 1.5 m between the camera and the treadmill is recommended
on the side of the treadmill to achieve optimum video recordings.
Preferably place a gym mat or other padding in the safety zone behind the treadmill in
order to soften falls.
The floor where the device is installed must be absolutely even and horizontal.
The belt of the treadmill must be checked after installation, or if the treadmill is moved to
another place, and adjusted if required. (Please refer to the user manual of the treadmill
manufacturer for this purpose.)
Set up the measuring system such that the socket for the mains connection is easily
accessible at all times and the device can be disconnected from the power supply.
Once the treadmill has been installed securely and is horizontalyl leveled, it can be con-
nected to a suitable mains socket and set into operation.