zebris Medical GmbH
Canid Gait Technical Specifications and User Manual
Page 9/33
3.5
Safety instructions for the Treadmill
The treadmill belonging to the CANID GAIT measuring system is a very powerful device.
For safe operation of the CANID GAIT system it is mandatory to adhere exactly to the
safety regulations described in the following.
The measuring procedure on the treadmill must never be commenced without a thorough
instruction of the patient by trained personnel. No measurements may be taken without a
supervisor.
Do not place the treadmill on an unstable ground.
Do not set up the system near a source of heating or in direct sunlight in front of a window
as a strong rise in temperature can lead to inaccurate measuring results.
Directly behind the treadmill it is mandatory for a safety zone of 2 m in length and 1 m in
width to be kept free,
and ought to be padded (with a soft mat). No items may be left in
this zone during operation (such as video camera, lighting equipment etc.).
Dangerous drawing-in gaps are located at the rear end of the running belt and along its
sides and (if existing) on the elevating mechanism. Do not wear any loose clothing that
could get caught up in the rollers. Make absolutely sure that if a person trips over, their
long hair, loose clothing, jewellery, etc., do not come into contact with the rear part of the
treadmill belt (e.g. wear a hair net). Due to danger of stumbling, do not place any clothing
or jewellery on or within close proximity of the treadmill.
During operation, the Emergency STOP facilities must always be within easy reach for
the user and the operating personnel.
Pull out the power plug before transporting the treadmill.
3.6
Prohibited Use
Improper and/or prohibited use of the measuring system is impermissible and zebris warn
explicitly against all prohibitions included in this section.
Do not try to service the treadmill in any manner other than that described in this user
manual. By the removal of the protective covers it is possible that you could expose your-
self to lethal high voltages or other hazards.
We also point out that if any changes are made to this certified device or its accessories
without the prior written consent of zebris, your legal right to operate the device will be
void. If changes are made to the device without obtaining approval, the operator is obli-
gated to carry out suitable investigations and tests in order to guarantee safe use.
The use of the measuring system under the influence of alcohol, drugs or narcotics is
strictly prohibited.
zebris measuring systems may not be operated in any other environmental conditions
than those listed in the section "Specifications", (e.g. in wet zones, moisture-prone areas,
or in climatic, vacuum, hyperbaric or decompression chambers, etc.). Direct contact with
liquids must always be avoided, as the measuring system is not protected against enter-
ing liquids. Liquids entering the device can cause fire, electric shock or other severe ac-
cidents.