figure3-23, zero point of coordinate in the upper left will increase one
correspondingly. Users can continue scanning in this interface, The
maximum distance can not be more than the maximal distance in the
first scan of this direction in second step.
Figure 3-23 Second scanning in X dire
ction
Figure 3-24 Second scanning in X direction
Sixth step
, after scanning between the second and third rebar in Y
direction, pressing
【
SAVE
】
key to end this scanning.
Seventh step
, similarly, scans for a second time in the central
position between the tested third and fourth rebar in Y direction (such
as measuring line b3 shown in figure 3-24) and so on.
Eighth step
, after second scanning in X direction, pressing
【
a
】
key
shifting to the other direction, in this case, it should be Y direction.
Ninth step
, principles of second scanning in Y direction are same as
X direction, second scanning between the second and third rebar in
previous scanning in X direction and between third and forth rebar in
order……until the blank space between the No. m-1 rebar and No. m
rebar. No. m rebar is the last rebar in first scanning of this direction as
shown in figure3-25.