Operation manual of ZBL-R800, ZBL-R630, ZBL-C310A
Hotline: (86)010-51290405 Fax: (86)010-51290406
33
in Y direction and the setting range is 10
~
99mm.
5)
Correction
—to set method of correction (please see in 3.4.1.1
for the definition of this parameter). It is divided into two
types—automatic and none, the default one after starting up is
none. automatic means that instrument will correct
automatically in accordance with testing thickness of cover
and rebar space. When entering the grid-scan interface next
time, the default method is the previous set one.
Selecting “Start Scan” in parameter setting interface, the set values
will be stored, and then enters into testing interface as shown in
figure 3-14.
Attention: while pressing
【
ENTER
】
, please hold sensor
in the air and get away from ferromagnet.
If pressing
【
BACK
】
key in testing interface, it can go back to
parameter setting as shown in figure3-13 and can reset parameters.
3.4.2.2 Measurement and data display
Figure 3-14 Grid-scan
The contents and definitions in the interface in figure 3-14 are as
follows:
1. Scrollbar
—show the relative distance between sensor and rebar,