Operation manual of ZBL-R800, ZBL-R630, ZBL-C310A
Hotline: (86)010-51290405 Fax: (86)010-51290406
43
Figure 3-28 Parameter setting interface of profile-scan
At that time, the system firstly detects the battery (see 3.4.1“General
-scan” in power detection part), and prompts when the instrument
battery is low. Figure 3-28 is the parameter setting interface, the
parameter setting option and methods of setting is the same as general
-scan and data correction with grid-scan-, please see 3.4.1.1 and
3.4.2.1 for specific definition and operation.
After parameter setting, please select “Start Scan” and press
【
ENTER
】
to store parameter and enter into the interface shown in
figure3-29.
Figure 3-29 Profile-scan interface Figure 3-30 Profile scanning
Attention: while pressing
【
ENTER
】
, please hold sensor
in the air and get away from ferromagnet.
The horizontal solid line in the central position of figure 3-29 is the
designed thickness value, two upper and lower dotted lines are
maximum and minimum of thickness of cover (the upper and lower
range is calculated according to the set object type(beam or floor)and
GB50204-2002 ).
When the instrument displays the current thickness value is 0, the