Section 6:  Programming for Odor Intensity Operation

Y Z   S y s t e m s   •   8 8 2 5   N .   S a m   H o u s t o n   P k w y   We s t   •   H o u s t o n ,  Te x a s   •   U S A  •   7 7 0 6 4   •   P :   2 8 1 . 3 6 2 . 6 5 0 0   •   F :   2 8 1 . 3 6 2 . 6 5 1 3

Conducting An Odor 

Intensity Test:

 (cont.)

Step 12:

 

Threshold Detection Level

 (TDL) Test: 

position nose within 3/4 inch of the sniff chamber 

and with valve closed sniff exhaust. If an unchar

-

acteristic odor is detected allow the instrument to 

operate for an additional 2 minutes. If uncharacter-

istic odor persists, perform an exhaust background 

evaluation test as specified in Section 7 - System 

Maintenance. Slowly open the flow valve and sniff 

exhaust. Continue to open the valve and sniff, 

getting breaths of fresh air between sniffs. Contin

-

ue the procedure until the first hint of a new odor 

is detected. This is the 

threshold detection level

Threshold, as used here, is the minimum Intensity 

of a gas in air when one detects a new or different 

odor (above the background/characteristic odor of 

instrument). Remove nose from sniff chamber.

Step 13:

 Press the 

Record Test Level

 (RTL) key 

recording the 

threshold detection level

*.

Step 14: 

Press the 

Test 

key to toggle your DTEX 

instrument into the Odor Intensity test mode. Quick-

ly open the flow valve until the RDL Flow value in 

the display equals the gas concentration in air % 

desired for this test.

Step 15: 

Position nose within 3/4 inch of the sniff 

chamber. Make a mental note of a rating for the 

intensity of the odor in the air/gas mixture, for use 

in step 18. 

*Typical ratings are (1) absent, (2) barely detect-

able, (3) RDL readily detectable, (4) strong, or (5) 

very strong or obnoxious. 

Step 16:

 Press the 

Record Test Level

 (RTL) key.

*refer ASTM D6273-98 definitions

Содержание DTEX DX1000G/L

Страница 1: ......

Страница 2: ......

Страница 3: ...Dual Process DTEX Instruction Operating Manual Version 05132020...

Страница 4: ......

Страница 5: ...onnections 7 Section 3 System Control Electronics 9 Display and Keypad 9 Special Keys 10 Section 4 Preparing Your DTEX for a Test 13 System Configuration 13 Section 5 Programming For Odor Concentratio...

Страница 6: ...YZ Systems 8825 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page VI DX1000G L ver 05132020...

Страница 7: ...is how the text will appear throughout the manual Items that require action for example the pressing of a key for programming the controller will feature the action item in sentence case Bold Text fol...

Страница 8: ...ion system You ve made a wise investment for your company DTEX is a small handheld instrument which assists the user in determining the odorant intensity of Natural Gas LPG with menu driven step by st...

Страница 9: ...n Level TDL Read ily Detectable Level RDL or of Desired Gas Concentration in Air The gas then passes through the electronic mass flow sensor and on to the mixing chamber where it is mixed with intake...

Страница 10: ...Page 4 DTEX 1000G L ver 05132020 Section 1 First Things To Know About The DTEX YZ Systems 8825 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Notes...

Страница 11: ...on Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page 5 DX1000G L ver 05132020 Section 2 System Installation Standard System Components The primary components of the DTEX DX 1000G L...

Страница 12: ...5 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page 6 DX1000G L ver 05132020 Section 2 System Installation System Flow Schematic Illustrated below are the major intern...

Страница 13: ...YZ Systems 8825 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page 7 DX1000G L ver 05132020 Section 2 System Installation Standard System Connections...

Страница 14: ...YZ Systems 8825 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page 8 DX1000G L ver 05132020 Section 2 System Installation Notes...

Страница 15: ...YZ Systems 8825 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page 9 DX 1000G L ver 05132020 Section 3 System Control Electronics System Electronics Display and Keypad...

Страница 16: ...the event of a technical service or support inquiry to the factory or for customer records Use the arrow keys to view all available information The following information is displayed Model No Version...

Страница 17: ...he details key allows for review of all information relative to the test previously selected by the review key Use the arrow keys to view all available infor mation The following information will be d...

Страница 18: ...YZ Systems 8825 N Sam Houston Pkwy West Houston Texas USA 77064 P 281 362 6500 F 281 362 6513 Page 12 DX 1000G L ver 05132020 Section 3 System Control Electronics Notes...

Страница 19: ...ng less user and location data input each time a test is conducted The following items can be uploaded into the system using reporter software User names User personal identification numbers PIN Test...

Страница 20: ...n the immedi ate area Should the user detect or suspect a leak prior to or during a test immediately shut off the source isola tion valve and seek the cause Avoid conducting tests in closed or confine...

Страница 21: ...ay the following Checking Hardware automatic system hardware check prior to a test Step 3 Press Enter to skip the purge cycle if de sired to review data repeat a test etc Press P to purge the unit bef...

Страница 22: ...Enter key on the keypad to begin data configuration for a test Step 5 Set User by using the arrow keys select the user name and press Enter Proceed to Step 6 If the correct name is not present select...

Страница 23: ...by typing in user 2 digit personal identification number and press Enter The main display will now be present Step 7 Press the Test button on the keypad Step 8 Set location using the arrow keys select...

Страница 24: ...t New Item and press Enter New City type in the name of the New City and press Enter Set State Province using the arrow keys select the state and press Enter If the correct State Province is not liste...

Страница 25: ...ys and press Enter Set to nearest 500 feet Step 10 Verify Data verify that all data entered for test location is correct by using arrow keys to scroll through data Press Enter to continue Step 11 Is D...

Страница 26: ...alve and sniff getting breaths of fresh air between sniffs Contin ue the procedure until the first hint of a new odor is detected This is the threshold detection level Threshold as used here is the mi...

Страница 27: ...Section 5 Programming for Odor Concentration Operation Conducting An Odor Concentration Test cont Step 16 Test Complete Close valve and press Enter Step 17 Add Any Notes type in any supplemental data...

Страница 28: ...ithout venting To be used after data review or data entry when no gas source connection has been made V TO VENT UNIT vent the system before shut down Recommended after each test to prevent undue satur...

Страница 29: ...he following Checking Hardware automatic system hardware check prior to a test Step 3 Press Enter to skip the purge cycle if de sired to review data repeat a test etc Press P to purge the unit before...

Страница 30: ...er key on the keypad to begin data configuration for a test Step 5 Set User by using the arrow keys select the user name and press Enter Proceed to Step 6 If the correct name is not present select Add...

Страница 31: ...yping in user 2 digit personal identification number and press Enter The main display will now be present Step 7 Press the Test button on the keypad Step 8 Set location using the arrow keys select the...

Страница 32: ...w Item and press Enter New City type in the name of the New City and press Enter Set State Province using the arrow keys select the state and press Enter If the correct State Province is not listed se...

Страница 33: ...nd press Enter Set to nearest 500 feet Step 10 Verify Data verify that all data entered for test location is correct by using arrow keys to scroll through data Press Enter to continue Step 11 Is Data...

Страница 34: ...l the first hint of a new odor is detected This is the threshold detection level Threshold as used here is the minimum Intensity of a gas in air when one detects a new or different odor above the back...

Страница 35: ...eration Conducting An Odor Intensity Test cont Step 17 Test Complete Close valve and press Enter Step 18 Enter Intensity Rating Notes Typical ratings are 1 absent 2 barely detectable 3 RDL readily det...

Страница 36: ...ut venting To be used after data review or data entry when no gas source connection has been made V TO VENT UNIT vent the system before shut down Recommended after each test to prevent undue saturatio...

Страница 37: ...not expose the unit to debris Damage Will Result Keep lid closed and gas inlet capped when not in use Keep air intake and sniff chamber clean 4 Clean the unit with a damp cloth Avoid cleaners which c...

Страница 38: ...ation and inspection of the DTEX system is required at least once every two years Annual re calibration is available if required by local P U C Company Policy etc The unit displays the next sched uled...

Страница 39: ......

Страница 40: ...ther expressed or implied including but not limited to implied warranties or merchantability and fitness for a particular purpose with respect to the product This limited warranty gives you specific l...

Отзывы: