Section 5: Programming for Odor Concentra
tion Operation
Conducting An Odor
Concentration Test:
(cont.)
Step 18:
press the
PWR
key.
The following will be displayed:
Press:
When finished with a test, the operator has the
option to:
PWR TO SHUT DOWN:
power down the
system without venting.
To be used after data review or data entry
when no gas source connection has been
made.
V TO VENT UNIT:
vent the system before
shut down.
Recommended after each test to prevent
undue saturation of inlet hose, system
components and residual smell in the unit.
To vent:
1. Close the source isolation valve.
2. Remove the gas inlet hose from the
source connection.
3. Press
V
to vent.
4. Open the flow valve fully.
5. Wait until prompted on screen.
6. Close flow valve.
7. Press
PWR
to shut down.
ESC TO ABORT:
to review or re-run the
previous test.
PWR
to shut down
V
to vent unit
ESC
to abort