DX-1000G/L ver.05132020
Section 5: Programming for Odor Concentra
tion Operation
Conducting An Odor
Concentration Test:
(cont.)
Step 12:
Threshold Detection Level
(TDL) Test:
position nose within 3/4 inch of the sniff chamber
and with valve closed sniff exhaust. If an unchar
-
acteristic odor is detected allow the instrument to
operate for an additional 2 minutes. If uncharacter-
istic odor persists, perform an exhaust background
evaluation test as specified in Section 7 - System
Maintenance. Slowly open the flow valve and sniff
exhaust. Continue to open the valve and sniff,
getting breaths of fresh air between sniffs. Contin
-
ue the procedure until the first hint of a new odor
is detected. This is the
threshold detection level
.
Threshold, as used here, is the minimum concen-
tration of a gas in air when one detects a new or
different odor (above the background/characteristic
odor of instrument). Remove nose from sniff cham
-
ber.
Step 13:
Press the
Record Test Level
(RTL) key
recording the
threshold detection level
*.
Step 14: Readily Detectable Level
(RDL) Gas
Odor Test: again position nose within 3/4 inch of
the sniff chamber. Continue opening flow valve
until readily detectable odor of gas is attained. RDL
determination should be where Natural Gas / LPG
odor can be positively identified as Natural Gas /
LPG gas. Remove nose from sniff chamber.
Step 15:
Press the
Record Test Level
(RTL) key.
*refer ASTM D6273-98 definitions