To adjust the ramp:

Highlight the ramp time, and press dial to see your current setting;
Press dial and rotate it to adjust your setting;
Press dial to save your change.

Ramp

45min

           

                          Picture 3-1

Home          

 

      Picture 3-2

Ramp

   

10.1.4 Humidity level:

The humidifier moistens the air and is designed to make therapy more comfortable. If you are 
getting a dry nose or mouth, turn up the humidity. If you are getting any moisture in your mask, 
turn down the humidity. You can set the humidity level between 0 to 6, where 0 means disable this 
function, 1 means the lowest level of the humidity and 6 represent the highest level of the humidity. 
If the humidity level is 1 to 6, it means that the heater plate is heating when the device is running. 
The temperature of each level on the heater plate is as following (accuracy is ±4

)

1 : 33

            2 : 35

            3 : 40

            4 : 45

            5 : 50

            6 : 55

This temperature has been tested with empty water tank. The time required (warm-up time) to reach 
the set temperature from a starting temperature of (23°C ± 2°C) is 10 minutes. 

To adjust the humidity level:       

•  Highlight the humidity level, press dial enter the humidity page;
•  Press the dial and rotate it choose humidity level ,then press dial again to save your change and 

back to home page;

•  You can change humidity level at any time during therapy. 

Humidity

3 Level

    

                                                                     

                            Picture 4-1: Home                                             Picture 4-2: humidity         

 

f

10.2 Function that patient must be used under the direction of doctor(Clinical menu):

The patient should use the device by the direction of physician.       

Mode

Max Pressure

Min Pressure

Initial Pressure

Therapy Pressur

FPS Level

APAP

20.0

15.0

15.0

15.0

2.0

Clinical Menu

Mode

Initial Pressure

Therapy Pressure

Pressure Unit

FPS Level

Ramp (minutes)

CPAP

4.0

10.0

cmH2O

2

30

Clinical Menu

             Picture 5-1

Clinical menu (CPAP)                     Picture 5-2

Clinical menu (APAP)

10.2.1 Set Mode ( CPAP or APAP )

•  At home page, press and hold START/STOP button and dial at the same time for three seconds, 

the clinical menu will be shown on screen (see picture 5-1).

•  Rotate the dial, highlight the “Mode”, and then select CPAP or APAP.
•  After mode setting, press dial to save your change.
•  Highlight the “Back”, press dial to back to home page.

10.2.2 Set Pressure 

•  At clinical page, choose “CPAP” mode, Highlight “Initial Pressure” or  “Therapy Pressure”, press 

and rotate dial to set suitable value. 

•  You can rotate the dial to the right or to the left to rise or decrease therapy pressure (each step 

is 0.5 cmH

2

O / hPa).

•  After setting therapy pressure, press dial to save your change.
•  Highlight the “Back”, press dial to back to home page.
•  You can set “Max Pressure”, “Min Pressure”, “Therapy Pressure” and “Initial Pressure” of APAP 

mode according to above method.

Notice:

When you try to make the Max pressure lower than Min pressure, or Initial pressure, the device will 
let these three pressures at same value automatically.

10.2.3 Set Pressure Unit

•  At clinical page, highlight “Pressure Unit”.
•  Rotate the dial to the right or left, set your pressure unit (cmH

2

O or hPa).

•  After setting pressure unit, highlight the “Back”, press dial to back to home page.

10.2.4 Set FPS Level:

•  At clinical page, Highlight “FPS Level”.
•  Rotate the dial to the right or left, set your level of FPS (0-3 level).
•  After setting level of FPS, highlight the “Back”, press dial to back to home page.

10.2.5 Set Ramp:

•  At clinical page, Highlight “Ramp”.
•  Rotate the dial to the right or left, set your ramp time (0-45minutes, step value is 5 minutes).
•  After setting Ramp, highlight the “Back”, press dial to back to home page.

Содержание YH-480A

Страница 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Страница 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Страница 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Страница 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Страница 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Страница 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Страница 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Страница 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Страница 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Страница 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Страница 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Страница 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Страница 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Страница 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Страница 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...

Страница 16: ......

Отзывы: