background image

03

04

• 

If you use the humidifier, always place the device on a level surface lower than your head to 
prevent the mask and heated tubing from filling with water. 

•  Leave the water tank to cool for ten minutes before handling to allow the water to cool and to 

make sure that the water tank is not too hot to touch. 

•  Make sure that the water tank is empty before transporting the device. 
•  The proper placement and positioning of the mask on the face is critical to the consistent 

operation of this equipment. 

•  Not intended for use with patient whose upper airways have been bypassed. 
•  The time required for the device to warm from the minimum storage temperature between uses 

until the device is ready for its intended use when the ambient temperature is 20

 is about 2 

hours. 

•  The time required for the device to cool from the maximum storage temperature between uses 

until the device is ready for its intended use when the ambient temperature is 20

 is about 2 

hours. 

• 

This device is not intended for children, and patients who are physically or mentally deficient 

can’t use the device without assistance or supervision. 

•  The patient is an intended operator at home.

2. Introduction

The YH-480A self-adjusting devices is indicated for the treatment of Obstructive Sleep Apnea (OSA) 
in patients weighing more than 66 lb(30 kg). It is intended for home and hospital/institutional use. 
The YH-480A consists of main device, water tank, power adapter and cable. Mask and heated 
tubing are purchased. 

3. Scope of application

BreathCare PAP is a kind of the device which provides positive airway pressure to patient’s airway. 
This therapy can make patient sleep better.

4. Contraindications and adverse effects 

 

f

4.1 Contraindications:

Positive airway pressure therapy may be contraindicated in some patients with the following 
preexisting conditions:

•  Severe bullous lung disease
•  Pneumothorax
•  pathologically low blood pressure
•  Dehydration
• 

Cerebrospinal fluid leak

•  Recent cranial surgery or trauma

 

f

4.2 Adverse effects:

You should report unusual chest pain, severe headache, or increased breathlessness to 
your prescribing physician. An acute upper respiratory tract infection may require temporary 
discontinuation of treatment. 
The following adverse reactions may arise during the course of therapy with the device: 

•  Drying of mouth, nose, or throat
•  Nosebleed
•  Bloating
•  Ear or sinus discomfort
•  Eye irritation
•  Skin rashes

5. Package table

Item

Quantity

Item

Quantity

Main device

1

Mask (with mask accessories)

1

Water tank

1

Air filter

2

Heated tubing

1

User manual

1

Power adapter and Cable(as part 
of the device)

1

Bag

1

6. Picture and explanation for product

START/STOP button

Screen

Dial

Water tank

Main device

Air filter

Heating wire 

connection

Air outlet of main 

device to water tank

Air outlet of 

main device

Heater plate

Air outlet

Power interface

Manufacturer reserved, no use

USB interface (for USB flash 

disk only)

Содержание YH-480A

Страница 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Страница 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Страница 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Страница 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Страница 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Страница 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Страница 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Страница 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Страница 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Страница 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Страница 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Страница 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Страница 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Страница 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Страница 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...

Страница 16: ......

Отзывы: