10.2.6 Set Humidity Level
• At clinical page, Highlight “Humidity Level”.
• Rotate the dial to the right or left, set your humidity level (0-6level, step value is 1 level).
• After setting humidity level, highlight the “Back”, press dial to back to home page.
10.2.7 Erase Data
• At clinical page, Highlight “Erase Data”.
• Press the dial to the page where you can choose OK or Cancel about erase data.
• If you choose OK, your therapy data will all be erased.
Caution:
During your therapy, you can go to clinical menu by pressing and holding dial and START/STOP
button at the same time for three seconds. Section 10.2 clinical menu tells the method to change all
the parameters.
11. Caring for your device
It is important that you regularly clean your device to make sure you receive optimal therapy. The
following sections will help you with disassembling, cleaning, checking, reassembling your device.
f
11.1 Disassembling
2
1
3
5
4
1. Pull the power connector out of device;
2. Press the snap on the heated tubing and gently pull it out of the machine;
3. Hold the water tank at the top and bottom, press it gently and pull it away from the device;
4. Open the water tank and pour out the left water;
5. Hold both the cuff of the heated tubing and the swivel of the mask, then gently pull apart.
Notice: Don't pull the threaded portion of the heated tubing.
f
11.2 Cleaning
WARNINGS!
• Regular cleaning of the device and its accessories is very important for the prevention of
respiratory infections.
• To avoid electric shock, always unplug the device before cleaning.
• Use mild soap that is nontoxic to humans.
• Before cleaning, check whether the device has been disconnected from the power supply, whether
the power cord has been unplugged, and whether the water tank of the device has cooled down.
Make sure the plate has cooled down to room temperature, so you do not get burned.
• Do not open or modify the device. There are no user serviceable parts inside. Repairs and
servicing should only be performed by an authorized service agent.
CAUTIONS!
• Overheating of the materials could lead to early fatigue of these materials.
• Do not clean the device with bleach, chlorine, ozone,UV light,aromatic solution, moisturizing
factor, antibacterial soaps, sesame oil, or any other methods outside of this manual.
•
Do not immerse the device in any fluids.
11.2.1 Cleaning the Water Tank
You can clean the water tank with a soft cloth which does not scratch the water tank (dip the soft
cloth in liquid soap if necessary), rinse it thoroughly, and then wipe it dry with a soft cloth.
WARNINGS!
• Emptying and cleaning the water tank daily will help prevent mold and bacteria growth.
• Allow the water in the chamber to cool down to room temperature before removing it from the
device.
CAUTIONS!
• Clean the water tank only after the water in it cools. Make sure that no water enters the device.
• After cleaning, rinse the water tank throughly in clean water to make sure that no soap residue is
left; then wipe it dry with a lint-free cloth, so as to prevent calcareous accumulations.
• Inspect the water tank for any leak or damage. Replace the water tank if any damage is present.
• It is recommended to do daily cleaning of the water chamber.
11.2.2 Cleaning the Enclosure
Wipe the surface of the device with a soft, slightly damp cloth.
CAUTIONS!
• The device can only be used after the enclosure is dry, so that no moisture enters the device.
• It is recommended to clean the enclosure once a week.