11

12

10.2.6 Set Humidity Level 

•  At clinical page, Highlight “Humidity Level”.
•  Rotate the dial to the right or left, set your humidity level (0-6level, step value is 1 level).
•  After setting humidity level, highlight the “Back”, press dial to back to home page.

10.2.7 Erase Data

•  At clinical page, Highlight “Erase Data”.
•  Press the dial to the page where you can choose OK or Cancel about erase data.
•  If you choose OK, your therapy data will all be erased.

Caution: 

During your therapy, you can go to clinical menu by pressing and holding dial and START/STOP 
button at the same time for three seconds. Section 10.2 clinical menu tells the method to change all 
the parameters. 

11. Caring for your device

It is important that you regularly clean your device to make sure you receive optimal therapy. The 
following sections will help you with disassembling, cleaning, checking, reassembling your device. 

 

f

11.1 Disassembling

2

1

3

5

4

1. Pull the power connector out of device;
2. Press the snap on the heated tubing and gently pull it out of the machine;
3. Hold the water tank at the top and bottom, press it gently and pull it away from the device;
4. Open the water tank and pour out the left water;
5. Hold both the cuff of the heated tubing and the swivel of the mask, then gently pull apart.

Notice: Don't pull the threaded portion of the heated tubing.

 

f

11.2 Cleaning 

WARNINGS!

•  Regular cleaning of the device and its accessories is very important for the prevention of 

respiratory infections.

•  To avoid electric shock, always unplug the device before cleaning.
•  Use mild soap that is nontoxic to humans.
•  Before cleaning, check whether the device has been disconnected from the power supply, whether 

the power cord has been unplugged, and whether the water tank of the device has cooled down. 
Make sure the plate has cooled down to room temperature, so you do not get burned.

•  Do not open or modify the device. There are no user serviceable parts inside. Repairs and 

servicing should only be performed by an authorized service agent.

CAUTIONS!

•  Overheating of the materials could lead to early fatigue of these materials.
•  Do not clean the device with bleach, chlorine, ozone,UV light,aromatic solution, moisturizing 

factor, antibacterial soaps, sesame oil, or any other methods outside of this manual.

• 

Do not immerse the device in any fluids.

11.2.1 Cleaning the Water Tank

You can clean the water tank with a soft cloth which does not scratch the water tank (dip the soft 
cloth in liquid soap if necessary), rinse it thoroughly, and then wipe it dry with a soft cloth.

WARNINGS!

•  Emptying and cleaning the water tank daily will help prevent mold and bacteria growth.
•  Allow the water in the chamber to cool down to room temperature before removing it from the 

device.

CAUTIONS!

•  Clean the water tank only after the water in it cools. Make sure that no water enters the device.
•  After cleaning, rinse the water tank throughly in clean water to make sure that no soap residue is 

left; then wipe it dry with a lint-free cloth, so as to prevent calcareous accumulations.

•  Inspect the water tank for any leak or damage. Replace the water tank if any damage is present.
•  It is recommended to do daily cleaning of the water chamber.

11.2.2 Cleaning the Enclosure

Wipe the surface of the device with a soft, slightly damp cloth.

CAUTIONS!

•  The device can only be used after the enclosure is dry, so that no moisture enters the device.
•  It is recommended to clean the enclosure once a week.

Содержание YH-480A

Страница 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Страница 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Страница 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Страница 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Страница 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Страница 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Страница 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Страница 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Страница 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Страница 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Страница 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Страница 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Страница 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Страница 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Страница 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...