•  When the level of the humidity is 0, the heating plate is not heating. Also if the device is not 

delivering gas pressure, the heating plate will not heat either.

 

f

9.2 Stop therapy

1. Remove your mask; 
2. Press the Start / Stop button or the therapy will stop automatically after about one minute when 

the Smart Start / Stop is enabled; 

3. To power off your device, please separate the plug from the electricity.

10. Function 

 

f

10.1 Function that patient can safely use

10.1.1 Information:

You can read summary of sleep report on information page. Following parameters will be displayed.

•  AHI:

 Indicates the number of apneas and hypopneas per hour.

•  Used Time (h):

 The time duration for the last therapy (h).

•  Treatment Time (h):

 The time of the patient wore the mask (h).

•  Average Pressure:

 The average pressure for the last therapy (cmH

2

O/ hPa).

•  Average Leakage (Lpm):

 Average air leak volume per minute for the last therapy (LPM).

•  Total Time (h):

 The total time after the last therapy time (h).

•  P90:

 The pressure for the 90 percent of the last therapy (cmH

2

O/hPa). 

•  Version:

 Software version number used for the device.

•  SN:

 The serial number of device. 

AHI

Used Time (h)

Treatment Time (h)

Average Pressure

Average Leakage (Lpm)

0.0

0.0

0.0

0.0

0.0

Info

                            Picture 1-1: Home 

 

 

    Picture 1-2: Information  

10.1.2 Setting:

Your device has been set up for your needs by your care provider, but you may find you want to 

make small adjustments to make your therapy more comfortable. Highlight setting and press the dial 
to see your current settings. From here, you can personalize your options.

•  Wisdom Humidifying:

 Enable this function, device can moistens the air automatically according 

to the environment condition.    

•  Heated Tube: 

You can set the heating level between 0 to 5, where 0 means this function is 

disabled, 1 is the lowest heating level, and 5 is the highest heating level. The heated tubing is 
used to heat the air, so that the treatment will become more comfortable. The heating level of the 

heated tubing (0-5 level) equivalent to the following temperature (accuracy ± 2

). (The following 

temperatures were measured with the humidification off and the wisdom humidification off)

      0 : OFF            1 : 18

            2 : 21

            3 : 23

            4 : 25

            5: 28

When the heated tubing is properly connected to the device, the therapy page will display the 
heating line icon  

  , as shown in the figure:

Therapy

10:06

7.0

cmH2O

3

0

0

SPO2(%)

PR(bpm)

5.0-15.0

?

?

APAP

Mode

Pressure

•  Sound Reminder :

 Enable this function, if abnormal air leakage occurs, the device will give a 

sound reminder.

•  Smart Start/Stop:

 When this function is enabled, therapy starts automatically when you breathe 

into your mask. When you remove your mask, it will stop automatically after 1minute.

•  FPS Level:

 When this function is enabled, you may find it easier to breathe out. This can help 

you get used to therapy. You can set the humidity level between 0 to 3, where 0 means this 
function is disabled, 1 is the lowest level of FPS, and 3 is the highest level of FPS. The higher 
FPS level means more comfortable expiration you will get.

•  Backlight (%): 

You can set the back light of the screen. The range of setting is 0-100%.

•  Date:

 You can adjust time shown on the screen, include year, month and date.

•  Time:

 You can adjust time shown on the screen, include hour and minute.

•  Clock Alarm:

 Enable this function, you can setting time of clock as well.

•  Language:

 You can choose English or Chinese.

•  Filter Remind:

 Enable this function, this can notice you when to check and change filter.

•  Restore Defaults:

 This function can restore the device parameters to the factory defaults.

Time

Clock Alarm

Language

Filter Reminder

Restore Defaults

10:06

OFF

English

4Week

>>Enter

         

                            Picture 2-1

Home                                            Picture 2-2

Setting 

10.1.3 Ramp:

Ramp time designed to make you more comfortable at the beginning of the therapy; you can adjust 
the ramp time from 0 to 45 minute with an increment of 5min.

Содержание YH-480A

Страница 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Страница 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Страница 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Страница 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Страница 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Страница 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Страница 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Страница 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Страница 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Страница 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Страница 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Страница 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Страница 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Страница 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Страница 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...

Страница 16: ......

Отзывы: