11.3 Checking:

You should check power adapter and cable, water tank, heated tubing and air filter regularly in case 

of any damage.
a. Check the power adapter and cable

•  Wipe the power adapter and cable with the dry compress if it is dirty.
•  Replace the power adapter and cable if it is broken.

b. Check the water tank

•  Replace the water tank if it is broken or cracked.
•  Replace the water tank if the seal is torn or cracked.

c. Check the heated tubing 

•  Replace the heated tubing if there are any holes, tears or cracks.

d. Check the air filter

• 

Please check the air filter every week and replace the air filter at least every four weeks.

• 

If you find some particle blocked in the air filter, replace it more often.

To replace the air filter

2

1

4

3

1. 

Open the air filter cover;

2. 

Remove the dirty air filter;

3. 

Place a new air filter onto the air filter cover;

4. 

Close the air filter cover. 

Caution: 

Make sure the air filter is fitted at all times to prevent water and dust from entering the device.

 

f

11.4 Reassembling:

After fished all the cleaning steps, reassemble all these parts together. When the water tank and 

heated tubing are dry, you can reassemble the parts. 

• 

Open the water tank and fill it with distilled room temperature water up to the maximum water 

level mark.

•  Close the water tank and insert it into the side of the device.
• 

Connect the heated tubing firmly to the air outlet located on the rear of the device.

• 

Connect the free end of the heated tubing firmly onto the assembled mask.

12. Therapy data

The device records your therapy data for you and your care provider so they can view and make 
changes to your therapy if required. 
The data is recorded in your therapy device and then transferred to your care provider via a USB 
Flash Disk.

USB Flash Disk

USB Connector

1. Open the cover at the left side of the device;
2. After the machine page is transferred to the home page, push the USB Flash Disk into the USB 

Connector;

3. Remove the USB Flash Disk from the device after it wrote success.

13. Traveling 

You can take your device with you wherever you go. Just keep the following in mind.

•  Use the travel bag provided to prevent damage to the device.
•  Empty the water tank and pack it separately in the travel bag.
•  Make sure you have the appropriate power cord for the region you are traveling to. For 

information on purchasing, contact your provider or Yuwell.

14. Trouble shooting

When your device has trouble, look for the following table to find solution. Contact your physician, 

your provider or Yuwell if you cannot solve your problem. Do not try to open the device.

Содержание YH-480A

Страница 1: ...ay Pressure Units Suzhou Yuyue Medical Technology Co Ltd No 9 Jinfeng Road Suzhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 web address www yuyue com cn IFU V 68 version 0...

Страница 2: ...uction 03 3 Scope of application 03 4 Contraindications and adverse effects 03 5 Package table 04 6 Picture and explanation for product 04 7 Explanation of button 05 8 Installation 05 9 Therapy 06 10...

Страница 3: ...ified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation Por...

Страница 4: ...al institutional use The YH 480A consists of main device water tank power adapter and cable Mask and heated tubing are purchased 3 Scope of application BreathCare PAP is a kind of the device which pro...

Страница 5: ...t Caution Please change the distilled water in the water tank every day Do not add the distilled water above the maximum water level mark in case the water may enter the heated tubing and the device A...

Страница 6: ...the highest heating level The heated tubing is used to heat the air so that the treatment will become more comfortable The heating level of the heated tubing 0 5 level equivalent to the following temp...

Страница 7: ...Initial Pressure Therapy Pressure Pressure Unit FPS Level Ramp minutes CPAP 4 0 10 0 cmH2O 2 30 Clinical Menu Picture 5 1 Clinical menu CPAP Picture 5 2 Clinical menu APAP 10 2 1 Set Mode CPAP or APAP...

Страница 8: ...e device has been disconnected from the power supply whether the power cord has been unplugged and whether the water tank of the device has cooled down Make sure the plate has cooled down to room temp...

Страница 9: ...d fill it with distilled room temperature water up to the maximum water level mark Close the water tank and insert it into the side of the device Connect the heated tubing firmly to the air outlet loc...

Страница 10: ...Press dial to turn on the screen light Check the connection of power ensure it connect with device firmly My water tank is leaking Water tank assemble incorrectly or it is broken Check the water asse...

Страница 11: ...g Heated tubing Plastic hose about 1 8 m Maximum volume of water tank 260 10 mL Air outlet 22 mm complies with ISO 5356 1 2015 Temperature Maximum temperature of heater plate 55 4 131 14 4 Cut out 11...

Страница 12: ...rning hot surface Class ii equipment Use by date Batch code Temperature limitation for storage and transport CN Country of manufacture china Date of manufacture is on the right side of the symbol or b...

Страница 13: ...y test Compliance level Electrostatic discharge ESD IEC 61000 4 2 8kV contact 15kV air Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines Surge IEC 61000 4 5 1kV differential mod...

Страница 14: ...The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 duty cycle square wave signal c As an alter...

Страница 15: ...any service organization that has not been expressly authorized by Yuwell to perform such repairs Any damage caused as accident Act of God or human factor Product which is not involved in Quality warr...

Страница 16: ......

Отзывы: