3.1.5 Velocity adjusting steps
3.1 Velocity adjusting steps
Attention 1: Measure the standard block before the calibration, to ensure that
the current setting of instrument parameters can measure the standard test
block correctly.
Attention 2: Probe zero calibration, one point calibration and two point
calibration are suitable for single echo mode, dual echo calibration is suitable
for dual echo mode.
3.2 Preset Other Specifications
Press MODE to enter the specification configuration interface, which
including many specification adjusting options like FILE NUMBER,
MEASURING MODE, VIEW MODE, PROBE SETUP, MINIMUM ALARM,
MAXIMUM ALARM, NORMAL THICKNESS, THE MINIMUM OF B
SCAN,
THE
MAXIMUM
OF
B
SCAN,
RECTIFICATION,
RECTIFICATION WAVEFORM,
RESOLUTION, UPDATE
RATE,
LANGUAGE, UNITS, AUTO POWER DOWN,ERASE ALL FILES AND
DEFAULT SETUP. See the following figure:
3.2 SPECIFICATION ADJUSTING STEPS
FILE NUMBER – Select the current file. Total 400 files, and each file could
save 252 thickness values or waveforms.
MEASUREMENT MODE - Single echo and dual-echo mode. Single-echo
mode is used for common measurement; dual-echo mode is used for through
coating function.
VIEW MODE: normal mode, difference mode and limit scanning mode.
PROBE SETUP: selectable many probe models, including: