3.4 A-scan Interface Mode

In this mode, users could view the current thickness value and the A-scan
waveform at the same time. The right side of the interface is the specification
adjusting area, in which the specifications can be adjusted and finally solve
various difficult and complicated thickness measuring applications to the
capacity.

3.8 A-SCAN MODE INTERFACE

3.8 -SCAN INTERFACE MODE

1

waveform display area

2

gate

3

material velocity

4

the

magnifying selected specification

5

measuring point( the first point of

intersection between the waveform and the gate)

6

the present thickness

value

7

the blank confines 8

the beginning coordinate on the screen 9

the

specification adjusting area

10

the selected specification 11

large digits

mode 12

specification selection

13

battery power

Attention: when the probe and the object are not complete coupled, the font in
the various interfaces are in GREEN, while if they coupled well, in WHITE
color and when the either the upper or lower limited are exceed, the font and
waveforms are displayed in RED color.
In the A-scan interface, press the bottom right button NEXT

to

highlight the value to be adjusted. Then press the direction keys to adjust the
values. Up and down keys are used for small increments, while left and right
for large increments.

GAIN

Adjust the amplification of echo signal by instrument to allow

manual increase or decrease in ldB. This function is very useful for the testing
of attenuation materials (like metal cast).

RANGE

Adjust the range of waveform that the screen displays. The

waveform can be compressed or spread visually and it’s invisible if the
display range is set incorrectly and the echo waveform is beyond the display
area, however, the testing value can also be read correctly.

DELAY

Adjust the starting position of the waveform on the screen. The

waveform will move horizontally adjusting this value, and it’s invisible if the
delay is set incorrectly and the echo waveform is beyond the display area,
however, the testing value can also be read correctly.

BLANK

Hide the unnecessary and useless clutter in front of the main waves.

The red line on the bottom of the screen shows the blank confines. The
adjusting blank confines are the present range confines. In the real testing,
wrong measurement due to the material may exist, such as near surface serious
corrosion, AL material, inside defects, uneven component, lamination
structure and so on. While adjusting the gain or gate can solve part of the
problem, but only blank function can avoid the mistake when those clutter
echoes are higher than the bottom echoes.

15

16

3.9 The waveform before blanking 3.10 omit the front noise waves by blanking

Содержание UM-5 Series

Страница 1: ...15 3 5 Real Cases Analysis of A Scan 17 3 6 Operation of B scan Interface 19 3 6 1 B scan Display 19 3 6 2 Introduction of B Scan 19 3 7 Dual echo Thru coating Mode 20 3 7 1 A scan Interface in Dual...

Страница 2: ...splay is available by showing the echoes which enables the more extensive measure range more precise result and very little measurement error and finally makes the measuring processes more controllabl...

Страница 3: ...LF FULL Display Mode Thickness Value Min Max capture DIFF RR A Scan B Scan V Path Correction Automatic compensate the non linearity of dual element probe Update Rate Selectable 4Hz 8Hz 16Hz Material V...

Страница 4: ...eat capacity data storage function Stores 100 000 thickness values 1000 waveforms only UM 5DL 15 Measure the net thickness of the workpiece through the coating layer only UM 5D UM 5DL 2 Keyboard Funct...

Страница 5: ...If the material velocity is known for example the velocity of steel is 5900m s The sound velocity can be setting manually 3 1 1 Probe zero calibration Note The calibration result is 4 00mm only under...

Страница 6: ...on adjusting options like FILE NUMBER MEASURING MODE VIEW MODE PROBE SETUP MINIMUM ALARM MAXIMUM ALARM NORMAL THICKNESS THE MINIMUM OF B SCAN THE MAXIMUM OF B SCAN RECTIFICATION RECTIFICATION WAVEFORM...

Страница 7: ...TIFICATION MODE RF full wave half half RF RF describes the complete echo waveform full wave indicates the half echo and the overturned half echo half means putting off the half echo and turn the half...

Страница 8: ...erence between measured and nominal thickness and reduction difference as a percentage of nominal thickness as well as the current measured and nominal thickness values Before using this mode presetti...

Страница 9: ...mall increments while left and right for large increments GAIN Adjust the amplification of echo signal by instrument to allow manual increase or decrease in ldB This function is very useful for the te...

Страница 10: ...echo brake through the gate to achieve the correct measurement 3 12 REAL CASE 1 2 There are some defects in the testing object and the gate locks the defect echoes As showed in the following figure th...

Страница 11: ...ng or paint layer on the surface of the workpiece it will bring considerable error to the measurement results UM 5D and UM 5DL have the echo echo measurement method which can accurately measure the ac...

Страница 12: ...kness value 7 delete the selected data 8 The current cell selected Whether it is in the interface of thickness value A scan or B scan the current measured thickness values will be stored for short pre...

Страница 13: ...t 5 2 Measuring Methods 1 SINGLE POINT MEASUREMENT Using the probe to measure a random point on the surface of the object the reading displayed is the thickness value 2 DOUBLE POINT MEASUREMENT Measur...

Страница 14: ...time the gauge will not be turned on after running out of power 6 2Considerations 1 Please be cautious of the zero block s getting rust as couplant will be spread on the surface of it when calibrating...

Страница 15: ...5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0 110 1700 Polyethylene...

Отзывы: