TC510 (standard probe)
TC550
(composite crystal probe)
PT-08(standard probe)
PT-06(small tube probe)
PT-04(miniature probe)
GT-12(high-temperature probe)
ZT-12 (cast iron probe)
NULL(Other probes)
MINIMUM ALARM: Set the minimum thickness alarm value, range of
0.15-635mm. The result will be displayed in red if the actual thickness is less
than the minimum value preset.
MAXIMUM ALARM: Set the maximum thickness alarm value, range of
0.15-635mm. The result will be displayed in red if the actual thickness is more
than the maximum value preset.
NORMAL THICKNESS: Set the normal thickness, range of 0.15-635mm.
The real concrete application will be introduced in the DIFFERENCE MODE.
MINIMUM OF B SCAN: Set the minimum thickness of the B scan.
MAXIMUM OF B SCAN: Set the maximum thickness of the B scan.
RECTIFICATION MODE: RF+, full wave, half -, half +, RF-. RF describes
the complete echo waveform; full wave indicates the half + echo and the
overturned half - echo; half – means putting off the half + echo and turn the
half – over to +; half + means putting off the half – echo and only left the half
+ echo. RF- indicates the phase waveform of RF+.
3.3 HALF-WAVE POSITIVE
RECTIFICATION WAVEFORM: outline mode and filled mode.
RESOLUTION: Set the decimal digits of the measurement result. Metric of
×
.
×
and
×
.
××
and imperial of
×
.
××
and
×
.
×××
.
UPDATE RATE: Update the rate of measurement result. Optional 4Hz, 8Hz
and 16Hz.
LANGUAGE:Chinese/English/German/French/Japanese
languages
are
available
UINTS: Selectable units of mm/inch.
AUTO POWER DOWN: The instrument automatically shuts down after a
certain period of time without operation, which can be 5 minutes, 10 minutes,
20 minutes or manual shut down only.
DELETE ALL FILES
–
Empty the thickness readings and waveform of all
files.
DEFAULT SET UP
–
Set all parameters to their factory default value.
11
12