3.3 Normal Mode

UM-5 series have three measuring interface: normal mode, A-scan interface,
B-scan interface. And there are three display modes of normal interface:
Thickness value mode, difference/rate-of-reduction measurement mode, MAX.
/MIN. measurement mode. Select in the “VIEW MODE” of
CONFIGURATION.

ATTENTION:

When the probe and the object are not completely coupled, the

font in the various interfaces are in GREEN, when properly coupled, they are
displayed in WHITE color and when the either the upper or lower limited are
exceeded, the front are displayed in RED color.
THICKNESS VALUE MODE: the default interface. This interface mainly
shows the present thickness value with large digits.

3.5 NORMAL MODE INTERFACE

1

the present thickness value 2

probe types, gain degree, single-echo, measuring

units 3

material velocity 4

battery power display 5

A

scan interface

DIFFERENCE/RATE-OF-REDUCTION MODE: The interface displays the
difference (difference between measured and nominal thickness) and reduction
(difference as a percentage of nominal thickness), as well as the current
measured and nominal thickness values. Before using this mode, presetting the
nominal thickness is needed. The method can be taken according to chapter
3.2

3.6 DIFFERENCE MODE INTERFACE

1

the difference between the nominal value and the measured value

2

the ratio between the difference and the nominal value 3

the currently

measured thickness value 4

the nominal value 5

difference signal

6

material velocity 7

battery power 8

A

scan interface 9 -- probe type,

gain degree, single echo, measurement unit

LIMITS VALUE SCANNING MODE: This mode allows the user to catch the
real-time maximum and minimum thicknesses when test the thickness of
material continuously. It shows the minimum and maximum values during the
testing as well as the currently thickness. Press the RESET relevant to get the
limits when measuring the thickness.

3.7 LIMITS VALUE MODE INTERFACE

1

the current measured value 2

the maximum value

3

the minimum value 4

unit

5

material velocity 6

battery power 7

A

scan interface 8

reset

13

14

Содержание UM-5 Series

Страница 1: ...15 3 5 Real Cases Analysis of A Scan 17 3 6 Operation of B scan Interface 19 3 6 1 B scan Display 19 3 6 2 Introduction of B Scan 19 3 7 Dual echo Thru coating Mode 20 3 7 1 A scan Interface in Dual...

Страница 2: ...splay is available by showing the echoes which enables the more extensive measure range more precise result and very little measurement error and finally makes the measuring processes more controllabl...

Страница 3: ...LF FULL Display Mode Thickness Value Min Max capture DIFF RR A Scan B Scan V Path Correction Automatic compensate the non linearity of dual element probe Update Rate Selectable 4Hz 8Hz 16Hz Material V...

Страница 4: ...eat capacity data storage function Stores 100 000 thickness values 1000 waveforms only UM 5DL 15 Measure the net thickness of the workpiece through the coating layer only UM 5D UM 5DL 2 Keyboard Funct...

Страница 5: ...If the material velocity is known for example the velocity of steel is 5900m s The sound velocity can be setting manually 3 1 1 Probe zero calibration Note The calibration result is 4 00mm only under...

Страница 6: ...on adjusting options like FILE NUMBER MEASURING MODE VIEW MODE PROBE SETUP MINIMUM ALARM MAXIMUM ALARM NORMAL THICKNESS THE MINIMUM OF B SCAN THE MAXIMUM OF B SCAN RECTIFICATION RECTIFICATION WAVEFORM...

Страница 7: ...TIFICATION MODE RF full wave half half RF RF describes the complete echo waveform full wave indicates the half echo and the overturned half echo half means putting off the half echo and turn the half...

Страница 8: ...erence between measured and nominal thickness and reduction difference as a percentage of nominal thickness as well as the current measured and nominal thickness values Before using this mode presetti...

Страница 9: ...mall increments while left and right for large increments GAIN Adjust the amplification of echo signal by instrument to allow manual increase or decrease in ldB This function is very useful for the te...

Страница 10: ...echo brake through the gate to achieve the correct measurement 3 12 REAL CASE 1 2 There are some defects in the testing object and the gate locks the defect echoes As showed in the following figure th...

Страница 11: ...ng or paint layer on the surface of the workpiece it will bring considerable error to the measurement results UM 5D and UM 5DL have the echo echo measurement method which can accurately measure the ac...

Страница 12: ...kness value 7 delete the selected data 8 The current cell selected Whether it is in the interface of thickness value A scan or B scan the current measured thickness values will be stored for short pre...

Страница 13: ...t 5 2 Measuring Methods 1 SINGLE POINT MEASUREMENT Using the probe to measure a random point on the surface of the object the reading displayed is the thickness value 2 DOUBLE POINT MEASUREMENT Measur...

Страница 14: ...time the gauge will not be turned on after running out of power 6 2Considerations 1 Please be cautious of the zero block s getting rust as couplant will be spread on the surface of it when calibrating...

Страница 15: ...5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0 067 1700 Platinum 0 130 3300 Plexiglass 0 110 1700 Polyethylene...

Отзывы: