2-15
3
"4, 5$5
Next we will show a step-by-step instrument programming procedure to illustrate the flexibility
you have with the 2700 SELECT. For illustration purposes, we will demonstrate initial setup for
simultaneous determination of dextrose and L-lactate on a dual channel 2700 SELECT.
NOTE: If you have a single channel unit, this example has more information than you need.
However, it will cover all pertinent information to set up a single channel unit.
This is a straight-forward setup utilizing one calibration solution to perform simultaneous
chemistries. Follow along by pressing keys on your 2700 SELECT. This exercise will be a good
learning tool. After completing the exercise, you will be directed to parts of the manual related
to your specific chemistry interests.
EXAMPLE: Set up the YSI 2700 SELECT to measure D-glucose (dextrose) and L-lactate,
simultaneously. Let us assume you have installed a YSI 2329 Lactate Membrane on the black
probe and a YSI 2365 Glucose Membrane on the white probe. YSI 2357 buffer has been
reconstituted and poured into the Buffer Bottle and YSI 2776 Standard (2.50 g/L dextrose and
0.50 g/L L-lactate) has been installed into the Calibrator Bottle position. The instrument has
been powered and the printer paper installed.
Let us also assume that you have your sample in a test tube and want to program the unit to run
the sample three times in succession (triplicate). You would also like to identify the sample by
an identification number to be printed with the result. To conserve printer paper you would like
the "brief" report version of sample and calibration results. As recommended for Initial Setup,
you will use all default measurement parameters.
In Section 2.8 you became familiar with the 6 categories in the Setup Menu and you were last
instructed to press MENU to display Main Menu. This is the level in the menu where we want
to start. If you have entered Run Mode or Standby Mode, see page 3-1 for instructions to exit
either mode.
At this time you may want to refer to Figure 5.2, Menu Flow Chart.
Main Menu:
Please select instrument mode
[RUN] [STANDBY] [MENU]
Press [MENU].
Select instrument function
1-Service 2-Setup 3-Diagnostic
Press [2] for Setup.
Select setup: 1-General 2-MeasParameter
3-RunMode 4-Report 5-PrntSetup 6-Default
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 2700D
Страница 50: ...3 10 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 96: ...5 26 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 104: ...6 8 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 118: ...7 14 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 144: ...9 18 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 154: ...D 2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 156: ...E 2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 160: ...F 4 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 162: ...G 2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 164: ...H 2 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...