9
9
9
9----5
5
5
5
•
POSSIBLE CAUSE: A sample has been injected into the sample chamber at the
improper time.
•
CORRECTIVE ACTION: Enter PROBE diagnostics and cycle the buffer pump.
•
•
POSSIBLE CAUSE: Pinched, plugged, leaking or disconnected tubing.
•
CORRECTIVE ACTION: Fix or replace tubing.
•
•
POSSIBLE CAUSE: Buffer pump not running.
•
CORRECTIVE ACTION: Enter PUMP diagnostics and verify buffer pump
performance.
•
•
POSSIBLE CAUSE: Buffer pump not performing properly.
•
CORRECTIVE ACTION: Replace tubing.
•
•
POSSIBLE CAUSE: Stir bar not present.
•
CORRECTIVE ACTION: Disassemble chamber and reinstall stir bar.
•
•
POSSIBLE CAUSE: Stir bar not spinning.
•
CORRECTIVE ACTION: Enter STIR diagnostics and check motor
performance.
•
•
POSSIBLE CAUSE: Failing enzyme membrane.
•
CORRECTIVE ACTION: Perform daily operational checks and replace if
necessary.
•