10. Appendix A–Typical Performance Data
Testing of the YSI Model 1500 SPORT vs. the YSI Model 2300 STAT suggests
the following typical performance:
Capillary Tube Injector, YSI Model 1502
Capillary Tube Injector, YSI Model 1502
Capillary Tube Injector, YSI Model 1502
Capillary Tube Injector, YSI Model 1502
2.0
6.0
10.0
14.0
18.0
22.0
26.0
30.0
YSI 1500 vs 2300
YSI 2300 ( mmol/l )
WHOLE BLOOD - CAPILLARY / 0.25% TX-100 - LACTATE
30.0
24.0
22.0
20.0
18.0
16.0
14.0
12.0
10.0
8.0
6.0
4.0
2.0
0.0
Y = 0.9973X - .0705
DATA
26.0
28.0
1500 SPORT
( mmol/l )
Total Blood Lactate - YSI 1515 cell lysing agent added to the buffer solution in
both instruments.
Concentration
Concentration
Concentration
Concentration
Pooled Est. Std. Dev.
Pooled Est. Std. Dev.
Pooled Est. Std. Dev.
Pooled Est. Std. Dev. Mean
Bias
Mean Bias
Mean Bias
Mean Bias
0-10 mmol/L
0.1 mmol/L
0.0 mmol/L
10-20 0.2
0.1
20-30 0.4
0.0