Check for Leaks
Check for Leaks
Check for Leaks
Check for Leaks
Examine the tubing for leaks and check the sample chamber for large air
bubbles, which may be caused by either loose connections or worn pump tubing.
Refer to Section 8.6 Tubing Replacement if the tubing needs to be replaced.
Clean up spills
Clean up spills
Clean up spills
Clean up spills
Spills should be cleaned up immediately to prevent corrosion and biohazard
buildup. Blood spills must be carefully and safely removed. If there is any
evidence of salt build-up around the sample chamber, disassemble and clean the
chamber as described in Section 8.3.
8.2 Preventative Maintenance
The YSI 1503 Preventative Maintenance Kit contains all the supplies necessary
to keep your SPORT operating properly. Perform all the maintenance procedures
in Section 8.3 through 8.7 every 6 months, or as required.
IMPORTANT:
IMPORTANT:
IMPORTANT:
IMPORTANT: The Waste bottles, both reagent bottles, the 500mL mixing
bottle and the 3cc Syringe provided in the YSI 1504 Starter Kit should be
cleaned every week with a disinfecting agent. The bottles and Syringe should be
rinsed completely with distilled water after cleaning to remove any disinfecting
residue, which would cause membrane failure if it got into the system.
8.3 Cleaning the Sample Chamber
Clean the sample chamber every month.
Unscrew the probes from their housings. If the membrane has passed the tests in
Section 6.7, you don't have to replace it, but you shouldn't let it dry out. If the
chamber is off the instrument for more than 15 minutes, the membrane should
be replaced.
Disconnect the supply and waste line tubing from the fittings on the sample
chamber manifold.
Remove the two screws securing the injection block to the top of the sample
chamber and move it to one side. Remove the 2 screws securing the sample
chamber to the case panel.
Remove the screw holding the manifold to the chamber. Be careful not to lose
the two O-rings that seal the connection. Replace them if necessary.
Be careful not to discard the small magnetic stir bar. An extra one is supplied in
your maintenance kit. Clean the chamber with the appropriate disinfecting
agent, then flush it with copious amounts of warm water, followed by a distilled
water rinse.
Содержание 1500 Sport
Страница 2: ......
Страница 8: ...1 1 1 1 2 2 2 2 ...
Страница 46: ...5 5 5 5 1 1 1 12 2 2 2 ...
Страница 68: ...6 6 6 6 22 22 22 22 ...
Страница 78: ...7 7 7 7 10 10 10 10 ...
Страница 86: ......
Страница 100: ...9 9 9 9 14 14 14 14 ...
Страница 114: ...12 12 12 12 4 4 4 4 ...
Страница 116: ...13 13 13 13 2 2 2 2 ...
Страница 122: ...14 14 14 14 6 6 6 6 ...
Страница 124: ...15 15 15 15 2 2 2 2 ...
Страница 126: ...16 16 16 16 2 2 2 2 ...
Страница 128: ...17 17 17 17 2 2 2 2 ...
Страница 132: ...18 18 18 18 4 4 4 4 ...