<8. Detailed Data Setting>
8-4
IM 11M13A01-04E
Switch operation
Display
>
∧
ENT
b
1
0
>
∧
ENT
CAL
>
∧
ENT
CAL
>
∧
ENT
SPAn
Y
>
∧
ENT
21.00
%
>
∧
ENT
OPEn/20.84
Output hold time
during calibration
>
∧
ENT
20.84
%
>
∧
ENT
ZEro
Y
>
∧
ENT
0.98
%
>
∧
ENT
OPEn/0.89
>
∧
ENT
0.89
%
>
∧
ENT
CALEnd
>
∧
ENT
b
10
>
∧
ENT
Basic panel
display
Figure 8.1
Definition of During Calibration
In a semi-automatic calibration, “during calibration” is the time, starting when a calibration instruction
is executed with an infrared switch or a contact input, to make a series of calibrations, until the preset
output stabilization time elapses.
In an automatic calibration, “during calibration” is the time, starting when automatic calibration is
carried out at the calibration start time, until the preset output stabilization time elapses.
(4) “Error” appears when Err-01 to Err-04 are being issued