<7. Startup>
7-7
IM 11M13A01-04E
7.4.5 Changing Set Values
(1) Selecting numeric values from among preset values
Switch operation
Display
Description
>
∧
ENT
0
The set value is displayed after the parameter code selection. An example of how
to select either 0, 1, or 2 as the set value is given below.
(The currently set value is 0.)
>
∧
ENT
1
Touch the [
∧
] key once to change the current value from 0 to 1.
>
∧
ENT
2
Touch the [
∧
] key again to change to the numeric value to 2.
>
∧
ENT
0
If you touch the [
∧
] key again, the numeric value will return to 0.
Continuously touch the [
∧
] key, and the numeric values will change continuously.
>
∧
ENT
C
01
Display the desired numeric value and touch the [ENT] key.
The display will then return to the parameter code selection
(2) Entering numeric values such as oxygen concentration values and factors
Switch operation
Display
Description
>
∧
ENT
0
0.0
The set value is displayed after the parameter code selection.
An example of entering “9.8” is given below. (The currently set value is 0.0)
>
∧
ENT
0
0
.0
Touch the [
>
] key to move the position of the digit that is flashing to the digit to
be changed. Continuously touch the [
>
] key, and the position of the digit that is
flashing will move continuously to the right.
>
∧
ENT
0
9
.0
Touch the [
∧
] key to set the numeric value to 9.
Continuously touch the [
∧
] key, and the numeric value will change in sequence
from 0 to 1 to 2 to 3 … to 8 to 9 and back to 0.
>
∧
ENT
09.
0
Touch the [
>
] key to move the position of the digit that is flashing to the right.
>
∧
ENT
09.
8
Touch the [
∧
] key to set the numeric value to 8.
>
∧
ENT
09.8
Where the correct numeric value is displayed, touch the [ENT] key.
>
∧
ENT
09.8
If you touch the [ENT] key again, the flashing stops and the current set value will
be in effect.
>
∧
ENT
C
11
Touch the [ENT] key once again to return to the parameter code selection display.
(3) If invalid numeric values are entered.
Switch operation
Display
Description
>
∧
ENT
98.0
If an invalid numeric value (beyond the input range specified) is entered,
“Err” will appear for two seconds after touching the [ENT] key.
>
∧
ENT
Err
>
∧
ENT
0
0.0
“Err” appears for two seconds, and the display returns to the first set value.
Re-enter the numeric value.
Содержание ZR202S
Страница 12: ...xi IM 11M13A01 04E Introduction ...
Страница 18: ...Blank Page ...
Страница 61: ...Blank Page ...
Страница 129: ...Blank Page ...
Страница 141: ...Blank Page ...
Страница 153: ......
Страница 155: ......
Страница 157: ......
Страница 159: ...Blank Page ...
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......