<10. Other Functions>
10-14
IM 11M13A01-04E
Output-related Items in Group C
Code
Item
Tuning
Engineering unit
Default setting
C01 Analog output
0 Oxygen concentration
Oxygen
concentration
1 Amount of moisture content
2 Mixed ratio
C03 Output mode
0 Linear
Linear
1 Logarithm
C04 Output hold during
warm-up
0 4 mA
4 mA
1 20 mA
2 Set value
C05
Output hold during
maintenance
0 Not held
Held output just
before maintenance
service
1 Held output just before
maintenance service
2 Set value remains held
C06 Output hold during
calibration
0 Not held
Held output just
before calibration
1 Held output just before calibration
2 Set value remains held
C07 Output hold during error
occurrence
0 Not held
Held output at a
preset value
1 Held output just before abnormal
state occurs
2 Set value remains held
C11 Min. oxygen concentration See Section 8.2.
%O
2
0%O
2
C12 Max. oxygen concentration See Section 8.2.
%O
2
25%O
2
C30 Output damping constant
0 to 255
Seconds
0 second
C31 Set value during warm-up
2.4 to 21.6
mA
4 mA
C32 Set value during
maintenance
2.4 to 21.6
mA
4 mA
C33 Set value during calibration 2.4 to 21.6
mA
4 mA
C34 Set value in abnormal state 2.4 to 21.6
mA
3.4 mA
Note: “C07” and “C34” is not displayed when option code “/C2” or “/C3” (NAMUR NE 43 compliant) is specified.
Alarm-related Items in Group D
Code
Item
Tuning
Engineering unit
Default setting
D01 Oxygen concentration
high-high alarm setpoint
0 to 100
%O
2
100%O
2
D02 Oxygen concentration
high alarm setpoint
0 to 100
%O
2
100%O
2
D03 Oxygen concentration
low alarm setpoint
0 to 100
%O
2
0%O
2
D04 Oxygen concentration
low-low alarm setpoint
0 to 100
%O
2
0%O
2
D30 Oxygen concentration
alarm hysteresis
0 to 9.9
%O
2
0.1%O
2
D33 Delayed alarm action
0 to 255
Seconds
3 seconds
D41 Oxygen concentration
high-high alarm detection
0 Not detected
Not detected
1 Detected
D42 Oxygen concentration
high alarm detection
0 Not detected
Not detected
1 Detected
D43 Oxygen concentration
low alarm detection
0 Not detected
Not detected
1 Detected
D44 Oxygen concentration
low-low alarm detection
0 Not detected
Not detected
1 Detected
Continue to the next page
Содержание ZR202S
Страница 12: ...xi IM 11M13A01 04E Introduction ...
Страница 18: ...Blank Page ...
Страница 61: ...Blank Page ...
Страница 129: ...Blank Page ...
Страница 141: ...Blank Page ...
Страница 153: ......
Страница 155: ......
Страница 157: ......
Страница 159: ...Blank Page ...
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......