<10. Other Functions>
10-2
IM 11M13A01-04E
Table 10.1
Parameter Codes for Detailed Operation Data
Code
Item
Engineering
unit
Code
Item
Engineering
unit
A00 Selection
of display
items
0 Oxygen concentration
A50 Span correction ratio 0
%
1 Oxygen analyzer (0.0)
A51 Span correction ratio 1
%
2 Oxygen analyzer (0.0)
A52 Span correction ratio 2
%
3 Analog output selected
A53 Span correction ratio 3
%
A01 Oxygen concentration
%O
2
A54 Span correction ratio 4
%
A02
A55 Span correction ratio 5
%
A03
A56 Span correction ratio 6
%
A04
A57 Span correction ratio 7
%
A05
A58 Span correction ratio 8
%
A06 Air ratio
A59 Span correction ratio 9
%
A07 Cell temperature
°C
A60 Zero correction ratio 0
%
A08 Cold junction temperature
°C
A61 Zero correction ratio 1
%
A09 Meas. gas temperature
°C
A62 Zero correction ratio 2
%
A10 Amount of water vapor in exhaust gas
%
A63 Zero correction ratio 3
%
A11 Cell voltage
mV
A64 Zero correction ratio 4
%
A12 TC voltage
mV
A65 Zero correction ratio 5
%
A15 Cold junction voltage
mV
A66 Zero correction ratio 6
%
A16 Current output
mA
A67 Zero correction ratio 7
%
A20 Cell response time
Seconds
A68 Zero correction ratio 8
%
A21 Cell internal resistance
Ω
A69 Zero correction ratio 9
%
A22 Cell robustness
A70 Calibration history 0
YY.MM.
DD/HH.MM
A23 Heater on-time ratio
%
A71 Calibration history 1
YY.MM.
DD/HH.MM
A24 Oxygen concentration (with time
constant)
%O
2
A72 Calibration history 2
YY.MM.
DD/HH.MM
A25
A73 Calibration history 3
YY.MM.
DD/HH.MM
A26
A74 Calibration history 4
YY.MM.
DD/HH.MM
A30 Maximum oxygen concentration
%O
2
A75 Calibration history 5
YY.MM.
DD/HH.MM
A31 Occurrence of maximum oxygen
concentration
YY.MM.
DD/HH.MM
A76 Calibration history 6
YY.MM.
DD/HH.MM
A32 Minimum oxygen concentration
%O
2
A77 Calibration history 7
YY.MM.
DD/HH.MM
A33 Occurrence of minimum oxygen
concentration
YY.MM.
DD/HH.MM
A78 Calibration history 8
YY.MM.
DD/HH.MM
A34 Average oxygen concentration
%O
2
A79 Calibration history 9
YY.MM.
DD/HH.MM
A35
A80 Time
YY.MM.
DD/HH.MM
A36
A90 Software revision
Note: The blank parameter codes above are not used in the oxygen analyzer.
Содержание ZR202S
Страница 12: ...xi IM 11M13A01 04E Introduction ...
Страница 18: ...Blank Page ...
Страница 61: ...Blank Page ...
Страница 129: ...Blank Page ...
Страница 141: ...Blank Page ...
Страница 153: ......
Страница 155: ......
Страница 157: ......
Страница 159: ...Blank Page ...
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......