![YOKOGAWA YTA510 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/yokogawa/yta510/yta510_user-manual_3838199062.webp)
<9. Parameter Summary>
9-4
IM 01C50E01-01EN
Object
ID
Attribute
ID
Label
Description
Default value Handling
4.
TRANSDUCER
block
1
Tag Description
Memo field available to write anything.
Transducer
W
2
Model
Indicates the model name of the transmitter.
---
R
3
Serial Number
Indicates the device number of the transmitter.
---
R
4
Out Switch
Indicates the direction of the burnout switch on the integral
indicator. (Burnout Low / Burnout High)
Burn out High R
5
Wireless Status
Indicates the wireless communication status
1. Indicates either the initial Idle status or Join status.
(Idle status / Join status)
2. Indicates whether Contract (Pub) has been established.
(Not established / Established)
3. Indicates whether Contract (R/W) has been established.
(Not established / Established)
1. Idle status
2. Not
established
3. Not
established
R
6
Display Selection Selects the LCD display
1. Indicates the display status of the AI1/AI2 PV value.
(Displayed / Not displayed)
Displayed
W
7
LCD Mode
Selects the LCD mode.
1. Indicates the On/Off mode.
2. Indicates the bar graph display
(Bar graph: Not displayed / Bar graph: Displayed.)
1. Off
2. bar graph
displayed
W
8
YTA Option
Indicates the optional function supported by the device. This
cannot be changed by the user..
1. Indicates the unit system used. (SI system / Other)
Specify when
ordering
R
9
Special Cmd
Special function parameter.
0. Initial value at the time of reading (None)
1. Squawk mode
2. Deep-sleep mode
To start from the deep-sleep mode, either remove and reinsert
the battery pack or use the provisioning device tool or the device
configuration tool via infrared communication.
0
W
11
Power Saving
Mode
Selects the power saving mode
0.Off (Disable)
1.On (Enable)
1
W
12
Mesurement Rate Indicates the measurement period of process values.
30 seconds
R
14
LCD Intermittetnt
Time
Set the time to turn off display on the LCD indicator.
Unit: second
60 seconds
W
5.
AI1 block
6.
AI2 block
1
Process Value
AI1/AI2 is a temperature output object.
Indicates the primary analog value (or corresponding process
value) and status used to execute this function.
Allows updating data by specifying this for the Concentrator
object.
1. Value: AI1/AI2 output value.
2. Status: Indicates the status of the AI1/AI2 output value.
1. ---
2. ---
W
2
Mode
A universal parameter to indicate the block’s operation status.
O/S, Auto, and Man can be selected.
1. Target: Specifies the AI1/AI2 mode.
2. Actual: Indicates the present AI1/AI2 mode.
3. Permitted: Indicates the mode that can be specified in
Target.
4. Normal: Indicates the AI1/AI2 normal status mode.
1. Target =
Auto
2. Actual =
Auto
3. Permitted
= Auto |
Manual |
OOS
4. Normal =
Auto
W
3
Concentrator OID
Indicates the Concentrator object value that corresponds to the
data update of the PV value.
---
R