Model YPP6200 Analog Input Rate/Totalizer
Instruction Manual
38
Setting the Decimal Point (
dEc pt
)
The decimal point may be set with up to five decimal places or with no
decimal point at all. The rate, total, and grand total decimal points are
independent.
Pressing the Right arrow moves the decimal point one place to the right
until no decimal point is displayed then it moves to the leftmost position.
Pressing the Up arrow moves the decimal point one place to the left.
Programming the Rate/Totalizer (
prog
)
It is
very important
to read the following information, before proceeding
to program the meter:
•
The meter is factory calibrated prior to shipment to read in milliamps
and volts depending on the input selection. The calibration
equipment is certified to NIST standards.
•
Use
the
Scale
menu to scale process inputs (e.g. 4-20 mA). A
calibrated signal source is not needed to scale the meter.
•
Use
the
Calibrate
menu to apply a signal from a calibrator or a
flowmeter.
The
Program
menu contains the following menus:
1. Scale without a signal source
2. Calibrate with a calibrated signal source
3. Total time base & conversion factor
4. Grand total time base & conversion factor
5. Total reset mode for total & grand total
Note: The Scale and Calibrate functions are exclusive of each other.
The meter uses the last function programmed. Only one of these
methods can be employed at a time. The Scale and Calibrate functions
can use up to 32 points (default is 2). The number of points should be
set in the Advanced menu under the
Multi-Point Linearization
(
Linear
)
menu selection prior to scaling and calibration of the meter,
see page 76 for details.
The process input may be calibrated or scaled to any display value within
the range of the meter.