16-2
IM WT5000-01EN
Storage Control
Store Mode (Store Mode)
You can choose from one of the following methods for starting and pausing storage.
• Manual (Manual)
When you tap Rec or press REC, numeric data is stored at the storage interval for the number of times
specified by the storage count.
• Real-Time Control (Real Time)
Storage starts after you tap Rec or press REC and the scheduled storage start time is reached. You can store
numeric data at the storage interval until the storage stop time (or just the storage count) is reached.
• Integration Synchronization (Integ Sync)
• Storage starts after you tap Rec or press REC and integration starts. You can store numeric data at the
storage interval until integration stops (or just until the storage count is reached) is reached.
• Storage continues even when integration is reset by the
. When the integration timer is
reset, the storage interval timer is also reset.
• When independent integration is enabled, storage cannot be performed in integration-synchronized
storage mode.
• Event Synchronization (Event)
Storage starts after you tap Rec or press REC, the measured data is updated, and a user-defined event
occurs. You can store numeric data each time the measured data is updated until the storage count is
reached.
• Single Shot (Single Shot)
Numeric data is stored whenever you tap Rec or press REC. You can store numeric data until the storage
count is reached.
16 Data Storage