19-5
IM WT5000-01EN
Environment Settings (Preference)
Frequency Display for Low Frequency Input (Freq Display at Low Frequency)
When the frequency of the input signal is lower than the frequency that this Instrument can measure, you can
choose to display the frequency as “0” or “Error.”
MTR Display for Low Frequency Pulse Input (MTR Display at Low Pulse Freq, option)
When the pulse frequency of the torque or speed input signal and auxiliary input (Aux) is lower than the
frequency that this Instrument can measure, you can choose to display the values of motor evaluation
measurement functions as “0” or “Error.”
Decimal Point Format for CSV Files(Decimal Point for CSV File)
When you save data in ASCII format (.csv), you can choose what type of decimal point to use and how to
separate the data.
• Period (Period): The decimal point is a period, and the separator is a comma.
• Comma (Comma): The decimal point is a comma, and the separator is a semicolon.
Rounding to Zero (Rounding to Zero)
• ON
When the voltage or current measurement is one of the following values with respect to the measurement
range, Urms, Umn, Urmn, Irms, Imn, Irmn, and other measurement functions based on these measurement
functions are displayed and output as zero. λ or Φ is displayed and output as error [Error].
• When the Crest Factor Is Set to CF3
When Urms, Uac, Irms, or Iac is 0.3% or less. When Umn, Urmn, Imn, or Irmn is 2% or less.
• When the Crest Factor Is Set to CF6 or CF6A
When Urms, Uac, Irms, or Iac is 0.6% or less. When Umn, Urmn, Imn, or Irmn is 4% or less.
• OFF
The measured values are displayed and output as they are.
19 Utility