19-7
IM WT5000-01EN
Remote Control (Remote Control)
Communication interface for controlling this Instrument from a PC. GP-IB, USB, and Network are the available
communication interfaces.
For details, see the Communication Interface User's Manual, IM WT5000-17EN.
• Only use one communication interface: GP-IB, USB, or Network. If you send commands simultaneously
from more than one communication interface, the instrument will not execute the commands properly.
• The REMOTE indicator on the front panel illuminates when the instrument is communicating with a PC in
remote mode. All keys except UTILITY (LOCAL) are disabled in Remote mode.
Network (VXI-11) (Network (VXI-11)
Connects this Instrument to a PC using Ethernet.
IP Address (IP Address)
Displays the TCP/IP setting that you specified in the Ethernet communication settings.
Timeout (Time Out)
If a connection cannot be established between the instrument and the PC within the amount of time specified
here, the instrument aborts the connection process.
You can set the timeout time to Infinity or a value between 1 and 3600.
You must set TCP/IP parameters to connect this Instrument to an Ethernet network.
Notes about Connections
• To connect the instrument to a PC, be sure to use straight cables and to connect through a hub or router.
Proper operation is not guaranteed for a one-to-one connection using a crossover cable.
• Use one of the following types of network cable that conforms to the transfer speed of your network.
A UTP (Unshielded Twisted-Pair) cable
An STP (Shielded Twisted-Pair) cable
19 Utility