Operations when Values Are Held
If you press HOLD when storage recording is in progress, the displayed values are held, and storage is
performed as indicated below.
• When the storage mode is set to Manual and the storage interval is 00:00:00, when the storage mode is set
to real-time control and the storage interval is 00:00:00, or when storage mode is set to event synchronization
The storage operation stops, but integration does not.
• When the storage mode is set to Single Shot
The held display values are stored. Even during integration, the held display values are stored.
• In all other cases
The held display values are stored. During integration, the values being measured are stored.
• Media access indicator only illuminates when this instrument creates files at the start of storage and closes
them at the completion or end of storage, but this instrument is actually always accessing the storage
device. During the period between when the files are created and when they are closed, do not remove the
USB memory or turn off the power. Doing so may damage the storage device or corrupt its data.
• When a point has no numeric data, NAN, OL, OF, ERROR, or a blank space will be stored instead. The
places that contain blank spaces are in measurement functions such as the ones for the 0th (DC) and 1st
harmonic orders of ΦU and ΦI, where the screen display is also blank.
• Functions ΔU1 to ΔPΣ are stored according to the delta computation type that you selected in the delta
computation settings.
• When this instrument automatically changes the range during storage, because measured data is not
updated, the storage operation is suspended and no measured data is stored. After this instrument finishes
changing the range, storage resumes.