10-8
IM WT5000-01EN
Matrix display (Matrix)
Item Number to Set (Item No.)
This is the same as setting the item number in the 4-, 8-, and 16-value displays.
Function (Function)
This is the same as setting the function in the 4-, 8-, and 16-value displays.
• In the matrix display, if you choose a measurement function that does not require an element or wiring unit
(such as η1 to η4, F1 to F20, Ev1 to Ev8, etc.), data is displayed in the first column.
• If the function (Function) is set to Speed, Torque, Pm, Slip, or SyncSp, the data is displayed as follows:
Column 1: Motor1 data
Column 2: Motor2 data
Column 3: Motor3 data
Column 4: Motor4 data
Harmonic Order (Order (k))
This is the same feature as the harmonic orders on the All display.
Resetting the Display Items (Reset Items)
This is the same as resetting the display items in the 4-, 8-, and 16-value displays.
Column Settings (Column Settings)
Number of Columns (Column Num)
Set the number of columns to 4 or 7.
Column Number (Column No.)
Select the number of the column that you want to configure.
Element (Element/Σ)
• You can select the element/wiring unit from the options below. The available options vary depending on the
installed elements.
None, Element1, Element2, Element3, Element4, Element5, Element6, Element7, ΣA, ΣB, ΣC
• If you select None, no measurement data is displayed in the selected column.
• If the selected wiring unit does not have any elements assigned to it, because there is no data, “-------” (no
data) is displayed. For example, if elements are assigned to ΣA but not to ΣB, the measurement function for
ΣB is displayed as “-------” (no data).
Resetting the Settings (Reset Items Exec)
Reset the column settings.