2-33
IM 760201-01E
Features
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
2.8 Trend, Bar Graph, and Vector Displays
Trends of each measurement function, bar graphs of each harmonic, and vectors of the
fundamental signals of each element (when using the harmonic measurement option)
can be displayed.
Trend Display
The trends of all measurement functions that are measured during normal measurement
and harmonic measurement are displayed.
Voltage (Urms1) trend
Voltage (U1) frequency
trend
Trend Display Data
When the sampling of waveform display data is off during normal measurement, the
numeric data of measurement functions that are determined at each data update interval
is P-P compressed for each display segment (raster) and made into trend display data.*
When the sampling of waveform display data is on during normal measurement, the
numeric data of measurement functions that are determined whenever a trigger occurs is
P-P compressed for each display segment (raster) and made into trend display data.*
* In some cases, P-P compression is not performed.
Scale Settings (See section 8.5 for operating instructions)
The WT500 can automatically set the values at the top and bottom of the screen based
on the maximum and minimum values of the trend display data (this is referred to as auto
scaling). The values at the top and bottom of the screen can also be set manually (manual
scaling).
Horizontal (Time) Axis (See section 8.6 for operating instructions)
The time per division can be set in the range of 3 s to 1 day.
Split Display and Window Assignment (See section 8.7 for operating
instructions)
Up to eight trends, T1 to T8, can be displayed. You can set trends T1 to T8 to any
measurement function for any element. In harmonic measurement, you can also set the
trends to harmonic orders.
You can also divide the screen into as many as four windows, and each trend whose
display is turned on will be assigned in numeric order to one of these windows.
Display Interpolation, Graticule, Scale Value Display, and Label Display (See
section 8.8 for operating instructions)
The settings specified for waveform display are used.
Содержание WT500