7-4
IM 760201-01E
7.4 Setting the Time Axis
Procedure
1.
Press
FORM
to display the Wave Form menu.
2.
Use the
cursor keys
to select Time/div.
3.
Press
SET
to display the time axis selection box.
4.
Use the
cursor keys
to select a time axis. When the scale value display is set to
ON (see section 7.8), the time at the left edge of the screen (which is fixed at zero
seconds) appears in the lower left of the screen, and the time at the right edge of
the screen appears in the lower right of the screen.
5.
Press
SET
to confirm the selection.
Explanation
For an explanation of the waveform display feature itself, see section 2.7.
The time axis is set in Time/div (time per grid division).
The time axis can be set up to the point in which the time corresponding to one screen
is equal to the data update interval, in 1, 2, 5 steps. For example, when the data update
interval is 500 ms, if you change the time-per-division in this order: 1 ms > 2 ms > 5 ms >
10 ms > 20 ms > 50 ms, the time corresponding to one screen changes in this order: 10
ms > 20 ms > 50 ms > 100 ms > 200 ms > 500 ms.
Time at the right edge of the screen
(amount of time contained in the screen)
Time at the left edge of the screen
(fixed at 0 seconds)
• The number of data points between
the left and right edges of the display.
• “P-P” indicates that the displayed
waveform has been P-P compressed
(see section 2.7 for details).
Time/div setting
Содержание WT500