12.3 Zero-Level Compensation
Procedure
Press
SHIFT+SET (CAL)
to execute zero-level compensation.
Explanation
Zero-Level Compensation
Zero-level compensation is the process of creating a zero-input condition using the
WT500 internal circuit and setting the level at that point to the zero level. It must be
performed to meet the specifications of this instrument (see chapter 14 for details).
• Pressing SHIFT+SET (CAL) executes zero level compensation.
• The WT500 automatically performs zero-level compensation after you change the
measurement range or input filter.
Note
• To make accurate measurements, we recommend that you execute zero-level compensation
after warming up the instrument for at least 30 minutes. Also, the ambient temperature
should be stable and within the specified range (see chapter 14 for details).
• If the measurement range and input filter remain the same for a long period of time, the zero
level may change due to the changes in the environment surrounding the WT500. If this
happens, we recommend that you execute zero-level compensation.
• There is a feature that automatically performs zero-level compensation during integration.
For details, see section 5.10.