5-28
IM 760201-01E
Integrating in Real-Time Integration Mode or Real-Time Continuous
Integration Mode
• Starting Integration
21.
Press
START/STOP
. The integration indicator changes to Ready (displayed in
yellow), and the WT500 is ready to integrate. When the scheduled integration
start time is reached, the integration indicator changes to Start (displayed in
green), and integration starts.
• Holding Integration
22.
Press
HOLD
. The HOLD key lights, and the numeric data display is held.
Integration continues.
• Un-Holding
23.
Press
HOLD
when the display is held. The HOLD key is no longer lighted, and
the numeric data display is updated. While the display is held, you can execute a
single measurement by pressing SHIFT+HOLD (SINGLE) to update the display.
• Stopping Integration
24.
Press
START/STOP
. The integration indicator changes to Stop (displayed in red).
The integration time and integrated value are held. If you press
START/STOP
before the scheduled integration end time is reached and then press
START/
STOP
again, integration will restart and continue until the scheduled integration
end time is reached.
In real-time integration mode, once the scheduled integration end time is reached, the
integration indicator switches from Start, displayed in green, to Stop, displayed in red. The
integration time and integrated value are held.
In real-time continuous integration mode, once the integration timer expires, the integration
time and integrated value are automatically reset, and integration continues until you press
START/STOP or until the scheduled integration end time is reached.
• Resetting Integration
25.
Press
SHIFT+START/STOP (RESET)
. The integration indicator changes to
Reset, the integration time and integrated value are reset, and their displays
change to “-----------” (no data).
5.9 Setting Real-Time Integration or Real-Time Continuous Integration