Calibration of the Harmonic Analysis Function
Connection
Use the same instruments as in case of AC power measurement and connect them in the same
way (refer to page 14-6 and 14-7).
Preparation
1 Set the voltage range of this instrument to 15V, and the current range to 1A.
2 Turn the harmonic analysis function ON.
Calibrating Currents
1 Set the ch1 of the synchronizer to 60Hz, ch 2 to 900Hz (15 times) and output these
frequencies.
2 Set the frequency of the voltage standard to 60Hz, the output voltage to 15V and output the
voltage.
3 Set the frequency of the current standard to 900Hz, the output current to 1A and output the
current.
4 Set the displayed number on display A of this instrument to 15.
5 Set the display function of display B to A and verify that the displayed value lies within the
specifications. In case of the WT130, verify each element 1, 2, and 3.
6 If required, change the ch2 setting of the synchronizer and the frequency of the current
standard, and verify another number.
7 Turn the output of the voltage and current standard OFF.
Calibrating Voltages
1 Set the ch1 of the synchronizer to 900Hz (15 times), ch 2 to 60Hz and output these
frequencies.
2 Set the frequency of the current standard to 60Hz, the output current to 1A and output the
current.
3 Set the frequency of the voltage standard to 900Hz, the output voltage to 15V and output the
voltage.
4 Set the displayed number on display A of this instrument to 15.
5 Set the display function of display B to V and verify that the displayed value lies within the
specifications. In case of the WT130, verify each element 1, 2, and 3.
6 If required, change the ch1 setting of the synchronizer and the frequency of the current
standard, and verify another number.
7 Turn the output of the voltage and current standard OFF.
Verification of Functions
Auto Range Operation
1 Set the voltage or current range of this instrument to Auto range. In case of no voltage or
current input, the voltage range will become 15V, and the current range will become 0.5A
automatically.
2 Press the V RANGE key to verify the 15V range and then press this key once more.
3 Press the A RANGE key to verify the 0.5A range and then press this key once more.
4 Connect the output terminal of the voltage standard (either AC or DC) to the voltage input
terminal of this instrument, and connect the current standard to the current input terminal.
5 Set the output voltage of the voltage standard to 600V and output this voltage.
6 Verify that the display shows “––––” as the measured voltage value for approx. 1.5 seconds
and then changes to 600V.
7 Turn the output of the voltage standard OFF.
8 Set the output current of the current standard to 20A and output this current.
9 Verify that the display shows “––––” as the measured current value for approx. 1.5 seconds
and then changes to 20A.
10 Turn the output of the current standard OFF.
14.2 Calibration