IM 253401-01E
4-4
4.3
Selecting the Measurement Range in case of
Direct Input
Relevant Keys
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
SAMPLE
V OVER
A OVER
MODE
RMS
V MEAN
DC
A
B
C
hour
hour
min
min
sec
V
VA
m
A
k
var
M W
TIME
V
PF
m
A
k
deg
M W
%
FUNCTION
AUTO
AUTO
MODE
1
Φ
3W
V
Hz
m
A
k
h
M W
TRIG
V RANGE A RANGE
HOLD
ENTER
INTEGRATOR
START
HARMONICS
MEMORY
INTEG SET
STOP
RESET
REMOTE
INTERFACE
OUTPUT
LOCAL
SETUP
h
SHIFT
WIRING
3
Φ
4W
3
Φ
3W
3V3A
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
*Shows the operation panel of the WT130. For the differences
between WT110 and WT130, refer to section 2.2, page 2-2, 2-3
Displays
relevant
keys and
indicator
Operating Procedure
(Display C)
1.
V RANGE
ENTER
3.
2.
ENTER
3.
(Display C)
1.
A RANGE
2.
End
End
• Voltage Range Setting
• Current Range Setting
• Perform operations following the thick line in the below menu.
• Press the ENTER key to confirm the selection or setting.
• When you want to leave the current menu during operation, press the key described
under step 1. The confirmed settings made until that point will be kept.
Explanation
Manual Range (fixed) versus Automatic Range (auto)
The measurement range can be of one of the following types. The initial setting is Auto range
ON.
• Manual range
Voltage range : selectable from 600/300/150/60/30/15V
Current range : selectable from 20/10/5/2/1/0.5A
• Auto range: Auto
The measuring range is adjusted automatically according to the input voltage or current as
follows. Overrange is handled the same way as for the manually selected range.