IM 253401-01E
App2-23
App
Appendix 2.3 Commands
Communication Commands 2
MEASure[:NORMal]:ITEM?
Function
Queries all settings related to the communication
output items of normal measured/computed data.
Syntax
MEASure[:NORMal]:ITEM?
Example
MEASURE:NORMAL:ITEM?
→
(Results are the same
as for MEASure:NORMal? )
MEASure[:NORMal]:ITEM:PRESet
Function
Sets the ON/OFF pattern for all communication
outputs of the normal measurement function.
Syntax
MEASure[:NORMal]:ITEM:PRESet {NORMal|
INTEGrate|CLEar}
Example
MEASURE:NORMAL:ITEM:PRESET NORMAL
Description
The following three patterns can be selected. The
same setting applies to the current all elements or
∑
.
NORMal : V/A/W
→
ON, others
→
OFF
INTEGrate : W/WH/AH/TIME
→
ON, others
→
OFF
CLEar :
all items
→
OFF
MEASure[:NORMal]:ITEM:{TIME|MATH}
Function Sets the communication output of
{integration
elasped time|MATH}
ON/OFF, queries about the
current setting.
Syntax
MEASure[:NORMal]:ITEM:{TIME|MATH} {<Boolean>}
MEASure[:NORMal]:ITEM:{TIME|MATH}?
Example
MEASURE:NORMAL:ITEM:TIME OFF
MEASURE:NORMAL:ITEM:TIME?
→
:MEASURE:NORMAL:
ITEM:TIME 0
MEASure[:NORMal]:ITEM:<normal
measurement function>?
Function
Queries communication output settings of the normal
measurement function.
Syntax
MEASure[:NORMal]:ITEM:<normal measurement
function>?
<normal measurement function>={V|A|W|VA|VAR|
PF|DEGRee|VHZ|AHZ|WH|WHP|WHM|AH|AHP|AHM|VPK|APK}
Example
MEASURE:NORMAL:ITEM:V?
→
:MEASURE:NORMAL:ITEM:
V:ELEMENT1 1;ELEMENT2 1;ELEMENT3 1;SIGMA 1
Description For the meanings of the symbols of functions, see
Note on page App2-11.
MEASure[:NORMal]:ITEM:<normal
measurement function>[:ALL]
Function
Sets the communication output concerning all
elements or
∑
ON/OFF at once.
Syntax
MEASure[:NORMal]:ITEM:<normal measurement
function>[:ALL] {<Boolean>}
Example
MEASURE:NORMAL:ITEM:V:ALL ON
MEASure[:NORMal]:ITEM:<normal
measurement function>:ELEMent<x>
Function
Sets the communication output concerning each
element ON/OFF, queries the current setting.
Syntax
MEASure[:NORMal]:ITEM:<normal measurement
function>:ELEMent<x> {<Boolean>}
MEASure[:NORMal]:ITEM:<normal measurement
function>:ELEMent<x>?
Example
MEASURE:NORMAL:ITEM:V:ELEMENT1 ON
MEASURE:NORMAL:ITEM:V:ELEMENT?
→
:MEASURE:
NORMAL:ITEM:V:ELEMENT1 1
MEASure[:NORMal]:ITEM:<normal
measurement function>:SIGMa
Function
Sets the communication output concerning
∑
ON/
OFF, queries the current setting.
Syntax
MEASure[:NORMal]:ITEM:<normal measurement
function>:SIGMa {<Boolean>}
MEASure[:NORMal]:ITEM:<normal measurement
function>:SIGMa?
Example
MEASURE:NORMAL:ITEM:V:SIGMA ON
MEASURE:NORMAL:ITEM:V:SIGMA?
→
: MEASURE:
NORMAL:ITEM:V:SIGMA 1
MEASure[:NORMal]:VALue?
Function
Queries normal measured/computed data set by
commands other than “
MEASure[:NORMal]:ITEM
”.
Syntax
MEASure[:NORMal]:VALue?
Example
MEASURE:NORMAL:VALUE?
→
10.04E+00,10.02E+00, 10.03E+00,49.41E+00,...
Description
• The renewal of normal measured/computed data
output here occures when bit0 (UPD) of the
condition register (refer to page App2-38) changes
from high to low. For more details, refer to 2.2.6.
• For the output format of normal measured/computed
data, refer to page App2-24.
• When the harmonic analysis function is ON,
harmonic analysis data will be returned.