IM 253401-01E
14-5
14
Adjustment, Calibration and
T
rouble-Shooting
DC voltage
standard
EUT
H
L
EXT
±
(A)
EXT
±
(A)
EXT
±
(A)
• External sensor input (equipped with option /EX2)
Calibration
Regarding the combination of voltage and current ranges, we recommend applying the
following.
• Test the current ranges with the voltage range set to 150V;
• Test the voltage ranges with the current range set to 5A.
Of course testing can be carried out using other combinations as well.
1 Set the voltage or current range of this instrument to the testing range.
2 Set the output voltage of the DC voltage standard to the rated range value of this instrument,
and output the voltage.
3 Set the output current of the DC current standard to the rated range value of this instrument,
and output the current.
4 Fine adjust the output setting value of the voltage standard so that the voltage or current value
displayed on this instrument shows the rated range.
5 Read the output voltage or current setting value of the voltage standard and treat this as the
reference.
6 Verify that the power factor value displayed on this instrument shows the rated value. The
product of the voltage setting value and the current setting value of the voltage standard is the
calibrated power factor value.
Note
Before carrying out the calibration described above, verify that this instrument performs within its accuracy
specifications.
• Set the output of the DC voltage standard to the rated range of this instrument, read this voltage or
current value on the display on this instrument and verify that this value lies within this instrument’s
accuracy.
• Set the output of the DC voltage standard to the rated range of this instrument, read the power factor
value on the display on this instrument and verify that this value lies within this instrument’s accuracy.
14.2 Calibration