3-5
IM 707281-01E
Software Operation
3
Setting the Output Timing of the Trigger Signal
The trigger operation panel shown below is used to set the output timing.
Set the output timing of the trigger signal
Trig tab
Specify at what position of the output waveform to output the trigger signal in the [Point]
entry box. The range is from 1 to the data length. For details regarding the data length,
see page 1-7.
Setting the Memory Partition, Time Base, and Arbitrary Waveform Data Loading
The Misc operation panel shown below is used to set the following settings.
Select the number of channels
to use
Select the block to be loaded
Set the slot/channel to be loaded
Set auto load
Select the time base
Set the sampling interval
Select memory partitions
Set the output start/end block
Misc tab
Set the data length
Selecting the time base
Select which of the following time bases to use during data output in the [Time Base] list
box.
Internal : Sample the waveform data using the sampling interval based on the
module’s internal clock
BUSCLK : Sample the waveform data by synchronizing to the time base signal of the
measuring station (CMNCLK)*
* For setting procedures, see section 4.6, “Setting the Trigger Source/Time Base
Source/Arming” in the WE7000 User’s Manual (IM707001-01E).
Setting the sampling interval
Set the interval when sampling the waveform data using the internal clock. When auto
load is used to load waveform data in wvf format, the sampling interval of the waveform
data is also loaded. However, this value can be changed by entering a value in the
[Sampling Interval] entry box. A value can be entered in the [Sampling Interval] entry box
only when the time base is set to internal clock. The range is 10
µ
s to 10 s. The
resolution is 10
µ
s. The default is 10
µ
s.
3.3 Setting the Output Condition in the AG Mode