Connecting the Output Signal Wire
For WE7281
The output signal wires are connected to the terminal block of the 4-CH, 100 kS/s D/A module.
Pin arrangement of the terminal block
The output terminal is a clamp type terminal, and the pin arrangement is as follows.
Pin Number
Signal
Pin Number
Signal
1
FG (Frame ground*)
11
CH3 L (CH3 negative)
2
N.C. (No connection)
12
N.C. (No connection)
3
CH1 H (CH1 positive)
13
CH4 H (CH4 positive)
4
CH1 L (CH1 negative)
14
CH4 L (CH4 negative)
5
N.C. (No connection)
15
N.C. (No connection)
6
CH2 H (CH2 positive)
16
FG (Frame ground*)
7
CH2 L (CH2 negative)
8
N.C. (No connection)
9
N.C. (No connection)
10
CH3 H (CH3 positive)
* Ground potential
Connecting the signal wires
CAUTION
• Do not apply a voltage exceeding the following levels, as it may damage the module.
Maximum common mode voltage
: 250 VACrms or
±
250 VDC between the
ground and the L terminal
Maximum voltage between channels : 250 VACrms or
±
250 VDC between
terminals of different channels
• Do not connect to the N.C. (no connection) terminals. Applying signal to them
may damage the module.
• Do not apply external voltage across the output terminals of the same channel,
as it may damage the module.
WARNING
• Make sure that the conducting parts of the output signal wire are not showing from
the opening (insertion point) of the terminal block. It can lead to electric shock.
The terminal block can be removed from the module by loosening the two attachment
screws as shown in the following figure. When removing or attaching the terminal block,
make sure to turn OFF the standby power switch on the measuring station.