1-12
IM 707281-01E
1.6 Waveform Generation in the FG Mode
Selecting the sweep mode
Select the sweep mode from the following choices.
Off
: Do not sweep
Repeat : Sweep continuously
Single : Sweep once when a trigger occurs
Trigger signal
Single&Hold : Perform a single sweep when a trigger occurs, then hold the last value and
output a continuous oscillation waveform
Trigger signal
Start sweep
Start sweep
Hold the
last value
Reset to
initial value
Setting the sweep time
Set the time duration for one sweep operation. The selectable range is 1 to 1000 s.
Note
• Sweep control will be used for trigger input signals if the sweep mode not set to OFF.
• Trigger outputs during a frequency sweep may differ from the appointed phase.