<6. TROUBLESHOOTING>
6-2
IM 12E01A06-02E
Table 6.1
Error Code List (2/3)
No.
Error
Occurrence
Problem
Return
E301 Turbidity zero
calibration
coefficient A
failure
During zero
calibration (zero
water).
Turbidity zero calibration coefficient A
newly determined after zero calibration
(zero water or lamp OFF) is outside of
the range of 0 to 5, or each coordinate
value of newly determined linearized
table is outside of the range of -99999 to
999999.
When problem is
eliminated. When
[YES]/[NO] key is
pressed, error code
display disappears, at
the same time during
failure, error is once
removed.
E302
Turbidity slope
SL failure
During span
calibration
(standard
solution)
Turbidity slope value SL newly
determined after span calibration
(standard solution) is outside of the
range of 25 to 200%, or each coordinate
value of newly determined linearized
table is outside of the range of -99999 to
999999.
E303 Turbidity check
block failure
During span
calibration
(check block)
Turbidity slope value SL newly
determined after span calibration (check
block) is outside of the range of 50 to
150%, or each coordinate value of newly
determined linearized table is outside of
the range of -99999 to 999999.
E304 Turbidity zero
correction
factor B failure
During zero shift
calibration
Turbidity zero correction factor B newly
determined after zero shift calibration is
outside of the range of -9 to 9.
E305
Turbidity
sensitivity
correction
factor K failure
During span
calibration
(sensitivity
correction)
Turbidity sensitivity correction factor K
newly determined after span calibration
(sensitivity correction) is outside of the
range of 0.25 to 4.
E306
Turbidity
reference
sensitivity SO
failure
During turbidity
reference
sensitivity
calibration
Turbidity reference sensitivity SO newly
set after span calibration (check block)
is outside of the tolerance of 0.0001 to
2000, or each coordinate value of newly
determined linearized table is outside of
the range of -99999 to 999999.
E307
Turbidity
response time
failure
All calibrations
Turbidity self-adjustment is not complete
before self- adjustment time has
elapsed.
E321 Communication
error
During
communication
with PC
Any communication error during
communication with PC.
When problem is
eliminated in the next
communication (only
one time is OK). When
[YES]/[NO] key is
pressed, error code
display disappears, at
the same time during
failure, error is once
removed. (excluding
RS signal during
communication with
calibration/cleaning
controller.)