<3. OPERATION>
3-5
IM 12E01A06-02E
Span calibration using standard solution (1/2)
F3-6E.ai
[YES]
[NO]
[NO]
[NO]
(3 times)
[MODE]
[YES]
[
]
[MODE]
[YES]
[*] at least second
[
]
[YES]
[ENT]
[
]
HOLD
*SERV
HOLD
HOLD
HOLD
*SETP
*SERV
B
C
16
*CODE
0
*CAL.TP
{*SETP} -> {*RANGE} -> {*HOLD} -> {*SERV}
To enter operation level
From zero calibration
Before performing span calibration using
standard solution, standard solution/check
block selection should be set to "0 (standard
solution)" in CODE 16 as shown below.
• Enter "16" using
[
]and [
] keys
• Enter "0" using
[
] key